Transport
Dräger X-plore 8000
23
3. Switch on the fan unit by pushing
the
button on the con-
trol panel.
After switching on, the device performs a self-test. Correct
malfunction if the device fails to work properly or warning
devices are triggered (see section 4 on page 20).
4. Cover open end of the breathing hose with hand.
The fan unit starts to run faster after approx. 5 seconds. Af-
ter approx. 20 seconds an alarm is triggered.
Have the fan unit checked if the fan speed remains un-
changed and no alarm is triggered.
5. If you wish, you can switch off the fan unit by pushing
the
button on the control panel again.
5.3.6
Replacing the O-ring at plug-in or bayonet-type
hose connector
1. Use the O-ring removal tool to lift the old O-ring out of the
groove.
2. Insert new O-ring in the provided groove.
6
Transport
Transport in the original packaging or in optionally available
transport box.
7
Storage
Storing the whole system:
z
Remove filter and rechargeable battery.
z
Dry the components in a container or cabinet. Store them
dry and clean and protect them from direct sunlight and
thermal radiation.
Storing rechargeable batteries:
z
Deeply discharged batteries may get damaged after pro-
longed storage. Charge them to 50 to 70 % prior to stor-
age.
z
If storage lasts for over 6 months, charge them in the
meantime.
z
Do not store rechargeable batteries for prolonged periods
outside the recommended temperature range. This might
reduce the remaining capacity and number of potential
charge cycles.
8
Disposal
9
Technical data
Overall system
Rechargeable batteries
Standard rechargeable battery
Long-life rechargeable battery
Charger
This product must not be disposed of as municipal
waste. It is therefore marked with the symbol on the
left.
The product can be returned to Dräger free of charge.
Please contact your national Dräger Sales Organisa-
tion or Dräger for more information.
Batteries and rechargeable batteries must not be dis-
posed of as municipal waste. They are therefore
marked with the symbol on the left. Collect batteries
and rechargeable batteries according to local regula-
tions and dispose of at battery collection centres.
Flow rate of respiratory protective
device/helmet/visor:
170/190/210 L/min
Flow rate of half/full face mask:
115/130/145 L/min
Rated period of service:
4 hours
with standard rechargea-
ble battery
8 hours
with long-life rechargea-
ble battery
Operating temperature:
1)
1) For the battery charger and rechargeable batteries, refer to the
separate information provided in this chapter.
For other components, refer to the corresponding Instructions for Use.
-10 °C to 60 °C
Operating/storage area humidity:
1)
95 % relative humidity
Storage temperature:
1)
-20 °C to 60 °C
Noise level:
approx. 64 dB(A)
International Protection Code:
IP 65
Operating temperature:
-10 °C to 60 °C
Operating/storage area humidity:
95 % relative humidity
Storage temperature:
-20 °C to 50 °C
Charging temperature:
0 °C to 50 °C
Charging time:
< 4 hours
Operational life time after a full
charge:
approx. 4 hours
1)
Rated voltage:
10.8 V
Rated capacity:
3.35 Ah
Output power:
36 Wh
Charging time:
< 4 hours
Operational life time after a full
charge:
approx. 8 hours
1)
1) Varies depending on the preset flow rate and the employed filter
and facepiece type
Rated voltage:
10.8 V
Rated capacity:
6.70 Ah
Output power:
72 Wh
Input voltage:
15 V
Input current:
4 A
Output voltage:
9 - 12.6 V
Output current:
4 A
International Protection Code:
IP 30
Operating temperature:
0 °C to 50 °C