Invasive blood pressure (iBP)
Start monitoring the invasive blood pressure
The invasive blood pressure can be measured on two
channels simultaneously. The settings and functions of
the channels are identical and are described only once
below.
●
Connect pressure transducer to the parameter box
(colour-coded grey).
●
Connect infusion system and catheter/canula to the
transducer. Eliminate any air bubbles in the catheter/
transducer system, as they can impair measurement.
●
Apply the pressure transducer level with the heart.
●
If necessary, wait until the pressure transducer has
warmed up.
●
Connect the transducer dome to the catheter/infusion
system and open it to atmospheric pressure.
Calibrating the sensors
On the system screen
●
Press the
»parameter«
softkey.
●
Select »
iBP 1
« or »
iBP 2
« with the rotary control and
press to confirm.
●
Position the cursor frame over
»Calibration«
with the
rotary knob and press to confirm. The system
enquires whether the infusion system has been
exposed to atmospheric pressure.
●
Open the sensor dome and press to confirm.
The clock symbol
u
appears during the calibration
process. A tick (
✔
) then appears behind the function.
●
Close the transducer dome and reopen the connec-
tion to the catheter/infusion system; the pressure
trace and pressure values are displayed.
Or on the parameter box:
●
Press the zero alignment key
≠
of the corres-
ponding sensor twice; a tick (
✔
) appears behind the
menu item on the system screen.
A question mark (?) appears after the menu item if
a fault occurs during zero alignment. »
CAL ?
« is
displayed instead of the measured values and the
LED in the
≠
key flashes.
Parameter box
Invasive blood pressure (iBP)
94
g
d
AUTO
Start
Stop
Punct.
Act.
→
0
←
→
0
←
11:59
00:00
IPPV alarm limits
standard screen 1- adults
Warning:
Caution:
Advisory:
2. ECG
II
1mV
iBP1
160
0
0
CO
2
40
30
PAW
20
0
33
6.0
freq 10 V
T
0.60
36.8 37.2
pulse def. 4 ST + 0.05
65
122/ 89
pulse 65 (100)
32 3
plat 26 mean 21
pulse 65
99
O
2
N
2
O ISO
Fi
29 70 0.8
Fet
26 68 0.6
ª
auto set
pat. alarm.
timer
start
config.
alarm
info
screen
config.
limits
auto set
vent. al.
para-
meter
Select a menu position. Please confirm !
Change the settings. Please confirm again !
06-12-98
etCO
2
mmHg
MV L/min
T
1
T
2
°
C
HR/Pulse (ECG)
1/min
iBP1 ART mmHg
Peak mbar PEEP
SpO
2
%
Fi CO
2
0
iBP 1
function
on off
location
AORTA
active
alarms
sys dia
graph
absol. morph.
amplitude
160
pulse
on off
calibrate
u
g
d
AUTO
Start
Stop
Punct.
Act.
→
0
←
→
0
←
15629278
15729238
15829278