Connecting the anaesthetic gas scavenging system
●
Hook the receiving system to the mount: position the
slots of the receiving system over the two holders and
slide down.
Connecting the transfer hoses
1
Pass the transfer hose from behind through the hole
in the Cicero EM and connect the hose to the scaven-
ging connector of the breathing system.
2
Connect the transfer hose to the connector on the
receiving system.
●
Make sure the connection is secure!
3
Make sure that the connector for the second transfer
hose is sealed with a screw plug.
Do not seal the openings of the receiving system.
Otherwise the breathing system may be drained!
Connect sample line
for sampling CO
2
/ O
2
and anaesthetic agent.
If a Y-piece with Luer lock is used:
4
Connect the sample line to the Luer lock of the
Y-piece.
5
Connect the sample line to the Luer lock of the
water trap.
If a Y-piece without Luer lock is used:
6
Insert T-piece with filter into the patient connection
of the Y-piece – filter facing upwards to prevent
congestion by droplets of liquid.
7
Connect the sample line to the Luer lock of the
filter in the T-piece.
5
Connect the sample line to the Luer lock of the water
trap.
Only use original sample lines, others may lead to
changes in the machine's technical specifications!
Reassembling machine
Reassembling the parts
129
7
6
FULL
5
D
D
1
1
2
3
4