Checking the NiBP pneumatic circuit for
leaks
Pull the supply hose of the sphygmomanometer cuff from
the twin hose and connect it to a pump-ball via a T piece.
The cuff (normal size) remains in the measuring circuit,
tightly rolled up. Then proceed as follows:
●
Open the venting valve in the measuring circuit.
Wait at least 30 seconds.
●
Press the »
config
« softkey.
●
Move the cursor frame horizontally across the screen
to »
calibration
« by turning the rotary control. Press
the rotary control to confirm the selection.
●
Move the cursor frame vertically to »
more
« by turning
the rotary control. Press the rotary control to confirm
the action.
●
Move the cursor frame vertically to »
NiBP test
« by
turning the rotary control. Press the rotary control to
confirm the selection.
●
The cuff pressure displayed on screen should be 0.
●
Wait 30 to 35 seconds, and then press the »
AUTO
«
key on the parameter box.
●
Increase the pressure to the test pressure value with
the pump-ball and wait at least 30 seconds for the
system to settle.
●
Measure the pressure drop.
If the overpressure relief valve responds, the para-
meter box will be vented automatically at 300 mmHg
±
5%. It will also be vented if the pressure falls below
15 mmHg. Repeat the test.
●
Quit the calibration mode by pressing the rotary
control after selecting »
Check NiBP
«. Deactivate the
»
AUTO
« key.
Measurement system testing
Detach the cuff supply hose from the twin hose and
connect to normal pressure instead.
●
Start the test mode as described above for
»
Checking the pneumatic circuit for leaks
«.
●
With the pump and venting valve at normal pressure,
the test pressure can now be set in the system.
●
Quit the test mode as described above for »
Checking
the pneumatic circuit for leaks
«.
135
Check operational readiness
11:59
00:00
IPPV alarm limits
standard screen 1- adults
Warning:
Caution:
Advisory:
para-
meter
alarm
info
screen
config.
limits
auto set
vent. al.
auto set
pat. alarm.
timer
start
config.
06-12-98
Check system configuration. Please confirm !
IPPV alarm limits
11:59
00:00
Anesth.Agent
selection
man. auto
gas
Isoflurane
The anesthetic
agent is automati-
cally selected by
agent detection.
Settings
mode
adult neon.
alarm sound
1 2 3 4 5 6 7 8 9
ECG pulse tone
0 1 2 3 4 5 6 7 8 9
ECG pulse tone
SpO
2
modulated on off
curve speed 12,5 25 50
mm/s
colours
call standard
Alarms
default
CO
2
on off
SpO
2
on off
parameter-
box
on off
HLM
on off
Pmax
on off
AGas Mix
on off
Calibration
iBP 1
iBP 2
more
?
?
06-12-98
13929278
14029278