Checking workstation with checklist
Manual tests:
After switching on, the apparatus runs through a self-test.
Prerequisites:
Machine is fully equipped -
Instructions for Use are familiar -
User has been instructed on using the machine -
Duration: approx. 5 minutes
Switching on:
1
Press the master switch.
(Combined master switch for gas and power)
The self-test starts
The user must first check:
On the ventilator:
– Indication of software version, all indicator lamps light
up for approx. 2 seconds. A single tone sounds.
– The message
»
Self-test
«
appears briefly in the display window.
The Ventilator display goes dark. No operations on the
ventilator are required during the self test.
On the screen:
– The internal program memories are tested. All LEDs
and display elements light up for approx. 2 seconds.
The LED in the standby key
E
remains lit.
– Two warning tones sound.
– A clock symbol
u
appears on the screen = duration
of self-test.
– The keys are still inactive.
Please wait!
Shortly afterwards:
– The checklist appears.
●
Indication in user advisory field:
Is the Cicero EM o.k. according to checklist?
Please confirm!
●
The user is consequently prompted to check the
points listed.
The equipment must be checked every time,
immediately before being used!
21
Technology for Life
Cicero EM
Version 4.0
06-12-98
Self-test
In case of emergency: press any ventilator function key for quick startup
D
Self-Test
Is the Cicero EM o.k. according to check-list ? Please confirm !
Check-list Cicero EM
Vapor handwheel set to zero
fill - level ok
less than 6 months since last inspection
correctly seated and locked
Safety-fill plug secured in place
Gas scavenger operational
Breathing system complete, locked in place
Soda lime no colour change
Emergency breathing bag present, functional
Central gas supply all pressures greater than 2.7 bar
Backup cylinders all pressures greater than 50 bar
Suction functional
Gas delivery O
2
, Air, N
2
O present
ORC N
2
O flow changes proportionally to O
2
O
2
-flush audible flow, breathing bag fills
cancel
test
11:59
00:00
To be verified by the user prior to each use
06-12-98
Gas/Power
ON / I
Press
I
O
D
1
Preparing for use