Patient alarm AutoSet
Alarm limits can also be set
automatically once the
patient is stable
.
The user can set a new limit range with new upper and
lower alarm limits by pressing the softkey »
auto set pat.
al.
«.
The alarm limits set in the »
limits
« menu are deleted and
cannot be automatically reactivated. The default alarm
limits remain unaffected by the Autoset and can be
reactivated at any time via the functions »
config.
«,
»
alarms
« and »
default
«.
The alarm limits for SpO
2
and FiO
2
also remain
unaffected.
AutoSet cannot set the lower alarm limits for the heart
rate below 40 beats per minute and below 90 mmHg for
systolic pressure values (50 mmHg for neonates).
Deactivated alarm limits (– –) are not activated by
auto set!
73
System screen functions
Softkeys
Patient alarm AutoSet
11:59
00:00
The alarm limits of haemodynamic parameters have been automatically adapted
to the actual measurement values !
IPPV alarm limits
standard screen 1- adults
Warning:
Caution:
Advisory:
timer
start
para-
meter
config.
alarm
info
screen
config.
limits
auto set
vent. al.
auto set
pat. alarm.
06-12-98
13329278
Parameter
new measured value
Unit
Remarks
tolerance …
upper
lower
ECG/Pulse HR/Pulse
+20.0
–20.0
beats per
minute
but not less than 40 beats per minute
ST
+0.05
–0.05
mV
NiBP/ART SYS
+20.0
–20.0
mmHg
but not less than 90 mmHg (50 mmHg for neonates)
DIA
+20.0
–20.0
mmHg
AORTA
SYS
+20.0
–20.0
mmHg
DIA
+20.0
–20.0
mmHg
ART pul.
SYS
+15.0
–15.0
mmHg
DIA
+5.0
–5.0
mmHg
MIT
+10.0
–10.0
mmHg
CVP
MIT
+5.0
–5.0
mmHg
ICP
MIT
+5.0
–5.0
mmHg
mmHg
SYS
+20.0
–20.0
mmHg
DIA
+20.0
–20.0
mmHg
MIT
+20.0
–20.0
mmHg
TEMP
+1.0
–1.0
°
C
etCO
2
+5.0
–5.0
mmHg
Fi AGAS
HAL
+0.5
–0.5
Vol. %
ENF
+0.5
–0.5
Vol. %
ISO
+0.5
–0.5
Vol. %
DES
+0.5
–0.5
Vol. %
SEV
+0.5
–0.5
Vol. %