61
System screen functions
Configuring the system screen in »Standby«
Configuring the general settings
The following can be switched on (»
yes
«) or off (»
no
«):
units
Indication of the units.
Temperature
display
Option to display the body temperatures in
the bottom right-hand corner of the screen
(below the softkeys).
This display configuration saves space for
a numerical module.
Acoustic
The volume and tones are defined here.
»
0
« equals »
off
«, while »
1
« is the lowest and »
9
« the
highest volume.
alarm sound
Volume of the alarm
pulse tone
Volume of the pulse tone
SpO
2
-modulated
ECG pulse tone
The ECG pulse tone can optionally
be modulated with the SpO
2
value.
tone sequence
Specifies the tone sequence.
Either ISO or Dräger standard.
List entry
A menu is displayed to allow the user to select at
what times (fixed intervals) or events (warning, caution
message, NiBP measurement) list entries are to be
made.
In addition, this menu defines the source of the blood
pressure value and whether EEG entries are required.
If EEG entries are selected, they replace the temperature
entries.
Screen configuration menu. Please confirm !
Standby / Configuration
11:59
00:00
alarms inactive !
Settings
mode
adult neon.
alarm limits
parameter
screen
acoustic
list entry
transport
RS 232 (Medibus)
record (printer)
basic config.
standard - screen 1
standard - screen 2
standard - screen 3
colour
curve speed
general settings
06-12-98
unit
yes no
temperature-
display yes no
Alarm sound and pulse tone configuration menu. Please confirm !
Standby / Configuration
11:59
00:00
alarms inactive !
Settings
mode
adult neon.
alarm limits
parameter
screen
acoustic
list entry
transport
RS 232 (Medibus)
record (printer)
basic config.
alarm sound
1 2 3 4 5 6 7 8 9
pulse tone
0
1 2 3 4 5 6 7 8 9
pulse tone
source
ECG Pleth.
tone sequence
DRÄGER Euro-Stand.
06-12-98
List configuration menu. Please confirm !
Standby / Configuration
11:59
00:00
alarms inactive !
Settings
mode
adult neon.
alarm limits
parameter
screen
acoustic
list entry
transport
RS 232 (Medibus)
record (printer)
basic config.
time interval 1 2 5 10 off
trigg. by warning on off
trigg. by caution on off
BP source iBP 1 iBP 2
pEEG on off
with every
NIBP
06-12-98
11429278
11529278
11629278