Measuring ranges
for graph display .......................: 20, 40, 80, 160 and 320 mmHg
Max. pressure signal .................: – 10 to 320 mmHg for undistorted display
Measured parameters ...............: Systolic, diastolic and mean pressure
Measuring range.......................: – 10 to 320 mmHg
Resolution ................................: 1
mmHg
Unit of measure ........................: mmHg
Site labels.................................: ART, AORTA, APulm, CVP, ICP, ?
adjustment of the measuring label, alarm limits and display configuration are
automatically controlled by the site label.
Oxygen saturation (SpO
2
)
Measurement of functional SpO
2
saturation.
Input.........................................: IEC type CF, floating, Colour-coded brown.
Cable type................................: without preamplifier.
Wavelengths.............................: 660 nm (red) and 920 nm (infrared)
Measuring range.......................: 0 to 105 %
Saturation resolution .................: 1%
Accuracy ..................................: Adults:
0 to 50 %
undefined
50 to 70 %
±
3 %
70 to 100 %
±
2 %
Neonates:
0 to 70 %
undefined
70 to 95 %
±
3 %
95 to 100 %
not specified
Accuracy of Durasensor ...........:
±
3 % (additional)
Measuring modes .....................: Normal:
measurement over 5 to 7 seconds
Fast:
2 to 3 seconds
Slow:
10 to 15 seconds
Off:
Module deactivated
Plethysmogram .........................: Automatic amplitude control.
Pulse rate .................................: Derived from plethysmogram; no pulse rate tone in slow mode. The pulse rate tone
is controlled by the SpO
2
measurement when programmed accordingly.
Pulse rate resolution .................: 1/min
Pulse rate accuracy ..................:
±
1/min
Body temperature measurement
Inputs .......................................: Floating, IEC type CF, colour-coded green.
Parameter.................................: Two temperature values (T1, T2); both are monitored by alarm limits.
Measuring ranges .....................: 0
°
C to 45
°
C;
Resolution ................................: 0.1
°
C.
Accuracy ..................................:
±
0.1
°
C.
Self-test....................................: A fault message is output when one of the automatically tested fault limits is
exceeded.
Technical data