System screen functions
»Standby« screen
In Standby mode, the Cicero EM is immediately ready for
operation. The
»Standby«
screen contains the following
softkeys:
– the leakage test,
– deleting the trend and list memory,
– system screen configuration.
If the device is left in Standby mode for longer than two
minutes, the screen is automatically switched off (dark).
It is automatically switched back on by any user input
(key, rotary control, etc.).
Running the leakage test:
The sample line and the sample gas return must be
connected correctly before starting the leakage test.
They are then included in the test as part of the overall
system. An open T-piece or open Luer lock can result in
major leakages!
1
Press either the softkey »
leakage test
« on the system
screen or the »
TEST
« key on the ventilator.
2
Display in dialogue field on ventilator:
IPPV leakage test
followed by:
IPPV leak = xx mL
The menu for the leakage test appears on the system
screen, followed after a few seconds by the current
leakage and compliance values.
The manual ventilation bag and its hose are not
included in the test!
The machine reverts to »
Standby
« after the test.
Delete trend; e.g. for a new patient:
Trend memory, list and Parameter Box memory (if used)
are deleted!
3
Press the softkey »
delete trend
« on the system
screen. Description on page 71.
The system enquires again whether the trend is really
to be deleted.
●
Press softkey »
delete
« to confirm.
The original screen is restored if the softkey
»
do not delete
« is pressed.
46
System screen functions
»Standby« screen
Standby
Ventilator
SW-Version 7.n
06-12-98
Select ventilation mode at the ventilator or switch on screen by
Data Manager
SW-Version 4.n
06-12-98
Parameterbox
SW-Version 2.n
06-12-98
config.
delete
trend
leakage
test
Cicero-EM
1
O
MAN
SPONT
IPPV
P
max
P
PCV
SIMV
PCV
V
T
f
IPPV
T
I
:T
E
f
IMV
T
IP
:T
I
Flow
PEEP
TEST
IPPV Leakage test
1
2
Standby
Select ventilation mode at the ventilator or switch on screen by
config.
leakage
test
Cicero-EM
1
delete
trend
Ventilator
SW-Version 7.n
06-12-98
Data Manager
SW-Version 4.n
06-12-98
Parameterbox
SW-Version 2.n
06-12-98
3
1
07229278
06129278
07329278