List screen
Certain measured values which have been automatically
or manually saved are displayed here for documentation
(to facilitate compilation of the anaesthetic record).
A time interval, delimited by the connection and
disconnection of the parameter box, is indicated in the list
by the label »
interval x start
« and »
interval x end
«. The
user programmed location of the device is also displayed
after the data has been transferred (see page 64).
The measured values and the warning or caution
messages triggering them are recorded line by line,
together with the time, either at specified intervals and/or
as a function of warning or caution messages or of an
NiBP measurement.
A list entry can be generated manually at any time by
pressing
U
indpendently of the screen in use.
Additionally a number of predefined labels can be
inserted into the record using the Softkeys.
For example:
»
Start of surgery
«
»
position change
«
»
Intubation
«
The list can run into several pages - so that you may have
to "leaf through" for a specific entry.
Return to previous page:
●
Select »
previous page
« with the rotary control and
press to confirm.
Continue to next page:
●
Select »
next page
« with the rotary control and press
to confirm.
Define list entry:
●
Select and confirm
»list entry«
. A menu is displayed
for the user to define the times (fixed intervals) or
events (warning, caution message, NiBP measure-
ment) at which list entries should be made.
●
In addition, this menu defines the source of the
displayed blood pressure measurement and whether
EEG entries are required. If EEG entries are selected,
they replace the temperature entries.
Print list entry:
●
Press the »
« softkey to print out the complete list
on a connected external printer.
70
System screen functions
System screen functions during operation
List screen
11:59
00:00
IPPV alarm limits
list screen - adults
Warning:
Caution:
Advisory:
II
1mV
time
11:00
11:05
11:10
11:13
11:15
HR/Pulse
1/min
64
67
63
62
63
NIBP
mmHg
Sys/M/Dia
iBP1/ART
120/100/90
123/103/88
iB P2/CVP
- - / 8/ - -
- - / 9/ - -
NiBP
128/107/93
SpO
2
%
90
97
96
90
97
CO
2
mmHg
i/et
1/35
1/37
1/31
1/34
1/33
PAW
mbar
Peak/Plat/PEEP
32/26/3
31/25/3
33/27/3
31/28/3
30/26/3
pulse def 4 ST + 0.05
65
MV
L/min
6.4
6.3
6.2
6.1
6.0
O2
Vol. %
i/et
30/26
31/25
32/24
31/25
33/23
anesth
agent
Fi/Fet
ISO
1.4/1.1
1.4/1.2
Enfl
1.2/1.1
1.2/1.1
1.1/1.0
temp
°
C
T
1
/T
2
36.8/37.2
36.9/37.3
36.7/37.1
36.8/37.1
36.9/37.3
start of
surgery
event
2
event
1
extuba-
tion
end of
surgery
position
change
intuba-
tion
start of
anaesthesia
12:30 Interval 1- start
12:45 63 120/100/90 90 1/35 6.4 30/26 1.4/1.1 32/26/3 60 36.8/37.2
12:50 62 122/102/89 97 1/37 6.3 31/25 1.4/1.2 31/25/3 60 36.9/37.3
13:00 Interval 1- end
list entry next page preceding page
Cpat
ml/mbar
60
60
60
60
60
Warning
start of anaesthesia
SpO
2
SpO
2
!!!
!!!
ª
I
HR/Pulse ECG
1/min
06-12-98
1/min
11:59
00:00
IPPV alarm limits
list screen - adults
Warning:
Caution:
Advisory:
II
1mV
pulse def 4 ST + 0.05
65
start of
surgery
event
2
event
1
extuba-
tion
end of
surgery
position
change
intuba-
tion
start of
anaesthesia
ª
I
list entry
next page preceding page
time interval 1 2 5 10 off
trigg. by warning on off
trigg. by caution on off
BP source iBP 1 iBP 2
pEEG on off
with every
NIBP
HR/Pulse ECG
06-12-98
13029278
13129278