Pressure relief ..........................: 90
±
10 mbar
Pressure limiting valve...............: Variable between 5 and 70 mbar,
±
15% of set value
Patient connections ..................: ISO cone 22 mm
;
Anaesthetic gas scavenging......: Nominal socket diameter 27 mm or ISO cone 30 mm
;
(
=
EN 1281-1)
Leakage ventilation system........: < 120 mL/min at 30 mbar
Maximum pressure in
breathing system ......................: 10 to 80 mbar
±
20 % or + 3 mbar, but at least 10 mbar over PEEP.
Expiratory resistance.................: 4 mbar at a flow of 60 L/min (conforming to EN 740).
Inspiratory resistance................: 4 mbar at a flow of 60 L/min (conforming to EN 740).
Non-return valves......................: Back-pressure of a wet valve < 1 mbar
Opening pressure of a wet non-return valve < 1 mbar
Ventilator
Drive gas consumption .............: 2 L/min O
2
or AIR in operation
0 L/min in »
Standby
« mode
Tidal volume (V
T
).......................: <20 mL
±
30 % or
±
6 mL (as from software version 7.2)
20 to 100 mL
±
10 % or
±
10 mL
100 to 1400 mL
±
5 % or
±
15 mL
Frequency (f
IPPV
).......................: 6 to 80 per minute
(as from software version 7.2)
6 to 60 per minute
(software version < 7.2)
Minute ventilation (
*
E
)..............: Up to 25 L/min
T
I
:T
E
.........................................: 1:3 to 2:1
±
5 %
T
IP
:T
I
........................................: 0 to 60 %
±
5 %
Inspiratory flow
*
I
max ..............: Max. 75 L/min
f
IMV
...........................................: 3 to 80 per minute
±
5 % (f
IMV
< f
IPPV
) (as from software version 7.2)
3 to 60 per minute
±
5 % (f
IMV
< f
IPPV
) (software version 7.2)
PEEP........................................: 0; 2 to 15 mbar
±
10 % or
±
2 mbar
(P
max
- PEEP > 5 mbar, as from software version 7.2)
0; 2 to 15 mbar
±
10 % or
±
2 mbar
(P
max
- PEEP > 10 mbar, software version 7.2)
Trigger pressure .......................: – 1 mbar
±
0,5 mbar (invariable)
Inspiration flow in PCV mode ....: 5 to 75 l/min (PCV ventilator)
Operating pressure in PCV mode : 10 to 70 mbar (PCV ventilator)
Gas delivery
Low-flow measuring tubes ........: (Calibrated for 20
°
C, 1013 mbar)
O
2
:
0.02 to 0.5 L/min
±
10 %
At minimum scale:
+ 20 %, – 10 %
At minimum scale:
±
5 %
O
2
:
0.55 to 10.0 L/min
±
10 %
At minimum scale:
+ 20 %, – 5 %
N
2
O:
0.02 to 0.5 L/min
±
10 %
At minimum scale:
– 20 %, + 10 %
At minimum scale:
±
5 %
N
2
O:
0.55 to 10.0 L/min
±
10 %
At minimum scale:
+ 20 %, – 5 %
AIR:
0.2 to 14.0 L/min
±
10 %
At minimum scale:
+ 15 %, – 5 %
Delivery output..........................: O
2
, N
2
O, AIR at least 9 L/min each
Technical data