74
System screen functions
Softkeys
Ventilation alarm AutoSet
Alarm Info
11:59
00:00
The alarm limits of ventilation parameters have been automatically adapted
to the actual ventilator settings !
IPPV alarm limits
standard screen 1- adults
Warning:
Caution:
Advisory:
timer
start
para-
meter
config.
alarm
info
screen
config.
limits
auto set
pat. alarm.
auto set
vent. al.
06-12-98
11:59
00:00
IPPV alarm limits
standard screen 1- adults
Warning:
xxxx
xxxy
xxxz
Caution:
xxxx
xxxy
xxxz
Advisory:
xxxw
xxxx
xxxy
xxxz
timer
start
para-
meter
config.
screen
config.
limits
auto set
pat. alarm.
auto set
vent. al.
xxxx
xxxy
xxxz
xxxx
xxxy
xxxz
zzzz
zzzx
zzzy
zzzj
zzzq
alarm
info
06-12-98
13429278
13529278
Ventilation alarm AutoSet
Prerequisite:
The ventilator must be in »
IPPV
«, »
SIMV
« or »
PCV
« mode.
The ventilation alarms can be automatically adjusted to
the current settings of the ventilator by pressing the
»
auto set vent. al.
« softkey.
The alarm limits previously set in the »
limits
« menu are
overwritten and cannot be reactivated!
The default alarm limits are not affected by Autoset and
can be reactivated at any time via the functions »
config.
«,
»
alarms
« and »
default
«.
Deactivated alarm limits (– –) are not activated by auto
set!
Parameter new measured value tolerance
Unit
upper
lower
P
aw
PEEP + (P
max
- PEEP) / 4
P
max
+ 10
mbar
AMV
V
T
* f * 0.6
V
T
* f * 1.4 L/min
but always: over 0.5
Alarm Info
The warning and caution fields each have space for
displaying three entries. The advisory message field has
space for four entries.
If further messages are present, a complete list of all
entries in order of priority can be displayed on the
system screen by pressing the softkey »
alarm info
«.
The list can be read as long as the key is pressed.