![Dräger Cicero EM Instructions For Use Manual Download Page 72](http://html1.mh-extra.com/html/drager/cicero-em/cicero-em_instructions-for-use-manual_2526452072.webp)
72
System screen functions
Softkeys
Limits
Softkeys
1
On the right-hand side, beside the screen, there is
an unmarked touch-sensitive keypad.
2
The function of these keys changes and is indicated
on the screen.
Only those keys which can be activated are indicated on
the screen.
Limits
The display comprises the designation of the measured
value, the value actually measured on the patient (large
numerals) and the set upper and lower alarm limits (small
numerals) under the symbol for the relevant alarm limit
(lower:
<
, upper:
>
).
A deactivated alarm limit is indicated by dashes in the
numbers field, and an alarm limit that cannot be activated
is indicated by a blank field.
The limit value settings only apply temporarily! They are
overwritten by the standard alarm limits when
– the Cicero EM is switched on.
– the system screen is switched to »
Standby
«.
– the setting »
Default
« is selected and confirmed for
»
Alarms
«.
– the mode is switched over between »
neon.
«/»
adult
«.
To change alarm limits:
●
Move the cursor frame to the required alarm limit with
the rotary control and press to confirm.
●
The value can then be altered by turning the rotary
control until the required value has been obtained.
●
The value is confirmed by pressing the rotary control -
this value now represents the active limit.
Each active alarm limit is shown as a dashed line in the
dynamic gas curve displayed on the screen. In this
example: CO
2
curve.
Refer to the section describing the »
Alarm concept
«
from page 78 onwards for information on the links
between certain alarms.
The limit value menu is automatically displayed whenever
an alarm limit has been violated.
(see »
Alarm concept
«, page 78)
11:59
00:00
06.12.98
1mV
iBP1
160
0
0
CO
2
40
30
PAW
20
0
10.2
29
MV l/min
99
SpO
2
%
Freq 12 V
T
0.60
∆
O
2 --
Fi O
2
Vol %
T2
37.5
°
C
T1
36.8
°
C
33
etCO
2
mmHg
1/min
33 10.2 19
et CO2 MV Peak
12 0.81 1
Freq. VT PEEP
72
Peak mbar PEEP
32 3
Plat 18 Mean 6
HR/Puls
72
50 120
Pulsdef.
4
5
NiBP sys
122
80 160
SpO
2
92
91 --
et CO
2
33
30 50
in CO
2
--
5
AMV
6.2
3.0 12.0
Fi O
2
29
20 --
Fi SEV
0.8
-- 3.4
PAW
9
8 40
iBP 1 sys
125
80 167
iBP 2 mit
31
0 8
ST
-0.05
±
0.10
T1
36.8
-- --
T2
37.5
-- --
IPPV alarm limits
standard screen 1- adults
Warning:
Caution:
Advisory:
1. ECG I
pulse def. 4 ST + 0.05
auto set
pat. alarm.
timer
start
alarm
info
auto set
vent. al.
Select the required limits and confirm !
HR/Pulse (ECG)
limits
79 99 --
/
--
HR/Pulse SpO2 Art (1)
1/min % mmHg
0
CO
2
40
1
2
<
>
Pulse
72
50
120
23729278
05829278
09529278