Cleaning counter-electrodes and contact pieces:
1. Unscrew the lower pressure pieces and
remove the respective contact pieces (if
required, use pliers).
2. Use the brush included in the supply to clean
the counter-electrodes with the same
detergent. After cleaning, flush the counter-
electrodes with water.
3. After cleaning the counter-electrodes, the
contact pieces must also be cleaned. For this,
use the tooth brush.
Blow all the parts dry and re-install the analyser
block in the reverse order.
NOTE:
The inclined surface at the shaft of the measuring
and calibration electrode must not block the
aperture to the flowmeter or to the mixing spiral.
The surface where the electrodes and contact
pieces are connected must be dry.
6.2 Exchanging diaphragm and
sealing rings
Dr. Co recommends to annually change
the diaphragm and all sealing rings at the analyser
block.
6.2.1 General hints
To exchange the diaphragm and the sealing rings,
the analyser block must be dismantled.
NOTE:
Carry out all works in a clean, dry place.
6.2.2 Means required
To dismantle the analyser block, you need:
a SW 7 socket spanner
a cross-tipped screwdriver
a special screwdriver for the diaphragm
(special tool)
an electrode removing tool (special tool)
a test tube brush (accessory kit)
Vaseline
15% KCl solution
a clean cloth as base
a mild cleaning solution (e.g. using
dishwashing agent)
a set of sealing rings DG 50.35.00
diaphragm DG 50.09.00
NOTE:
In order to maintain the analyser block, we offer
the service set DG 50.91.00.
It contains all required tools and auxiliary
substances as well as a service manual.