4.3.2 Carrying out the calibration
1. Press
the key. The calibration starts after
10 seconds and takes 60 seconds.
2. When the 60-second countdown starts, lead
the hose from the sample return line into the
measuring cylinder and collect the sample.
3. Calibration is completed after 60 seconds.
Remove the hose from the measuring
cylinder.
Then press the
key to confirm the
message.
4. Read and enter the volume of the liquid
collected. Press
or
to select a number.
Press
to mark the next digit.
5. Use to confirm your input.
Figure4.3
WARNING!
The value to be entered should be between 150ml
and 180ml.
If a wrong value is entered, for example 500ml,
the Kf factor will be re-adjusted. This can only be
rectified by the Dr. Co maintenance
service.
6. The Kf factor is displayed for 3 seconds.
Then the display returns to the calibration
menu.
7. Press
to return to the "Measuring" screen.
4.3.3 Check Kf factor
1. Press to open the main menu.
2. Press to mark the menu item STATUS.
3. Press
. The submenu STATUS displays
the K
f
factor.
Figure: 4.4