Figure: 4.19
6. After successful calibration, the calculated Ks
factor is displayed for three seconds.
Then the
Digox 6
automatically shifts to the
"Measuring" screen.
4.6.2 Ref max calibration
1. Press . The calibration menu is opened.
2. Press
to select the menu item
REF MAX CALIBRATION and confirm with
.
3. In the following window press
and start
the calibration.
Figure: 4.20
4. The calibration starts. A countdown appears,
beginning with 50 seconds. After the
calibration, a reference concentration is
polled.
Use the cursor keys to enter it and confirm
with
.
5. After successful calibration, the calculated Ks
factor is displayed for three seconds.
Then the
Digox 6
automatically shifts to the
"Measuring" screen.
NOTE:
If the calibration curve is determined again, the
most recent point determined, the Ref min value is
used as a basis.
With Ref min = 0, the calibration curve crosses the
zero point.