17
Figure:3.2
NOTE:
When screwing in the reference electrode, excess
KCl solution will leave the bore (3) (Figure 3.2).
Please collect the liquid with absorbent paper.
WARNING!
The analyser block should now rest for about 15
minutes so that the diaphragm can take up the
KCl solution.
3.2 Connections
A flexible hose and two hose clips are included in
the scope of supply to connect the
Digox 6
.
NOTE:
Keep the feed line from the samplings point to the
analyser as short as possible.
If reductions, clutches or other hose adapters are
used, air bubbles could settle on it. Because of
this, the running-in time of the analyser is
prolonged in order to reach low concentrations.
WARNING!
Only use oxygen-proof hoses. Silicon hoses will
lead to higher oxygen values.
3.3 Taking the unit into service
Shortly press
to switch the
Digox 6
on.
The starting screen is displayed.
The
Digox 6
will at first perform a system check.
Then the message
"The current user of this unit is" ************
Version X.X“.
Then the "Measuring" screen appears (see section
2.1).
NOTE:
The message "Version X.X“ indicates the current
firmware version of the Digox 6.
To switch the
Digox 6
off, press
and hold it for
at least 5 seconds.