Example:
Calculator with a connected mechanical flow sensor (here: WESAN WP 222)
Hydrometer Reed Switch
Fig. Z
WESAN WP 222 with reed switch
6.1.3. Connecting temperature sensor
NOTE
The calculator is operated with separately approved pairs of Pt 100 or Pt 500
temperature sensors. The type of sensor to be used is printed on the front of the
calculator. Ensure that the approved temperature range of the temperature sensors
is the same as the temperature range of the calculator!
Handle the temperature sensors carefully!
The sensor cables are fitted with coloured type labels:
-
Red: sensor in hot line
-
Blue: sensor in cold line
Switch(Reed)
Calculator terminal
Pulse (white)
10
GND (brown)
11 (-)