58
Multimat
Easy
9 Error messages and troubleshooting
Display shows
Problem
Possible cause
Potential remedy
Nothing (stays dark)
Unit is not turned on or no mains
power. Automatic circuit breaker
has been activated.
Turn on the unit or fix the cause of the
power outage
.
Turn the unit off and on
F 05
Vacuum still prevails
No air is entering the system
Notify service technician
F 06
Bus error
Internal system error
If error recurs, notify service technician
F 07
Com error
Error in motor communication
Notify service technician
F 08
Excessive controller temperature
a) Firing support plate above the
controller is missing
b) Extremely high standby tempera-
ture while the chamber is open
Disconnect from mains and allow to cool for
about five minutes
a) Put support plate in place
b) Keep furnace closed when not in use
Restart furnace normally once cooled
F 09
Defective heating circuit
Heating muffle or control relay
defective
Notify service technician
F 10
Excessive temperature
Current temperature exceeds the
pre-set temperature by 35 °C or
more
Notify service technician
F 11
Reference error
Reference sensor defective
Notify service technician
F 12
Selected vacuum level not attained
Firing chamber gasket or vacuum
system leak
Notify service technician
F 13
Rapid cooling not completed
Stand-by temperature has not been
reached
Wait until the stand-by temperature has
been reached or terminate rapid cooling by
pressing ESC
F 14
Power outage has occurred
Short-term power outage during
firing
Not possible – check firing result
F 15
Heat sensor error
Internal measurement error
If error recurs, notify service technician
F 16
Battery error
Backup battery voltage too low
Battery to be changed by service technician
F 19
Access refused
Writing or reading access denied
for this file
Safe data in another file or under different
name
F 20
Out of memory
The unit’s internal memory is full
Delete any data or programs no longer
needed
F 05
Various failure states and errors are shown in the display in
the form of error messages, which starts with the letter
F
followed by
two digits
. The following table lists and
explains these messages: