background image

DAC-U2530EN-AG rev E

DAC-U2530EN-AG rev E

24

25

For more detailed information, refer to the User Manual (at www.defi btech.com).  

WARNINGS AND CAUTIONS

!

WARNINGS:

Immediate hazards that will result in

serious personal injury or death

.

•  Hazardous electrical output. This equipment is for use only by 

qualifi ed personnel.

•  Possible fi re or explosion.  Do not use in the presence of 

fl ammable gases or anesthetics. Use care when operating this 
device close to oxygen sources (such as bag-valve-mask devices 
or ventilator tubing).  Turn off gas source or move source away 
from patient during defi brillation, if necessary.

•  The DDU-2000 Series AED has not been evaluated or approved 

for use in hazardous locations as defi ned in the National Electric 
Code standard.  In compliance with IEC classifi cation, the DDU-
2000 Series AED is not to be used in the presence of fl ammable 
substance/air mixtures.

Conditions, hazards, or unsafe practices that 

may result in serious personal injury or death.

•  Not intended to be used in an environment with high-frequency 

electrosurgical equipment.

•  Improper use can cause injury. Use the DDU-2000 Series AED only 

as instructed in the User Manual and Operating Guide. The DDU-
2000 Series AED delivers electrical energy that can potentially 
cause death or injury if it is used or discharged improperly.

•  Improper maintenance can cause the DDU-2000 Series AED 

not to function. Maintain the DDU-2000 Series AED only as 
described in the User Manual and Operating Guide. The AED 
contains no user serviceable parts – do not take the unit apart.

• No modifi cation of this equipment is allowed.
•   Electrical Shock Hazard. Dangerous high voltages and currents 

are present. Do not open unit, remove cover (or back), or attempt 
repair. There are no user serviceable components in the DDU-
2000 Series AED. Refer servicing to qualifi ed service personnel.

•   Lithium metal battery packs are not rechargeable. Any attempt 

to recharge a lithium metal battery pack may result in fi re or 
explosion. Do not attempt to recharge the primary battery pack.

•  Do not immerse battery pack in water or other liquids. Immersion 

in fl uids may result in fi re or explosion.

•  Do not attempt to recharge, short-circuit, puncture, or deform 

battery. Do not expose battery to temperatures above 50°C 
(122°F). Remove battery when depleted.

•  Do not let fl uids get into the DDU-2000 Series AED. Avoid spilling 

fl uids on the AED or its accessories. Spilling fl uids into the DDU-
2000 Series AED may damage it or cause a fi re or shock hazard. 

WARNINGS

 (continued)

•  Do not sterilize the DDU-2000 Series AED or its accessories.
•  Use only Defi btech disposable self-adhesive defi brillation pads, 

battery packs, and other accessories supplied by Defi btech or its 
authorized distributors. Substitution of non-Defi btech approved 
accessories may cause the device to perform improperly.

•  Do not open sealed pads package until pads are to be used. 

The packaging should be opened only immediately prior to use, 
otherwise the pads may dry out and become non-functional. 

•  Do not touch the patient during defi brillation. Defi brillation current 

can cause operator or bystander injury.

•  Do not allow pads to touch metal objects or equipment in 

contact with the patient. Do not touch equipment connected to 
the patient during defi brillation. Disconnect all non-defi brillator 
proof equipment from the patient before defi brillation to prevent 
electrical shock hazard and potential damage to that equipment.

•  Do not shock with defi brillation pads touching each other. Do not 

shock with gel surface exposed.

•  Do not allow defi brillation pads to touch each other, or to touch 

other ECG electrodes, lead wires, dressings, transdermal 
patches, etc. Such contact can cause electrical arcing and patient 
skin burns during defi brillation and may divert defi brillating energy 
away from the heart. 

•  The defi brillation pads are intended for one time use only and 

must be discarded after use. Reuse can lead to potential cross 
infection, improper performance of the device, inadequate 
delivery of therapy and/or injury to the patient or operator. 

•  Avoid contact between parts of the patient’s body and conductive 

fl uids such as water, gel, blood or saline, and metal objects, 
which may provide unwanted pathways for defi brillating current.

•  Do not connect the DDU-2000 Series AED to a PC or other 

device (using the USB port) while the unit’s electrodes are still 
connected to the patient.

•  Aggressive or prolonged CPR to a patient with defi brillation pads 

attached can cause damage to the pads. Replace the defi brillation 
pads if they become damaged during use.

•  Possible Radio Frequency (RF) interference from RF devices 

such as cellular phones and two-way radios can cause improper 
AED operation. Normally using a cell phone near the AED 
should not cause a problem; however, a distance of 2 meters 
(6 feet) between RF devices and the DDU-2000 Series AED is 
recommended.

•  CPR during analysis can cause incorrect or delayed diagnosis by 

the patient analysis system. 

•  Do not place adult defi brillation pads in the anterior-posterior 

(front-back) position.  A shock or no shock decision may be 
inappropriately advised.  The DDU-2000 Series AED requires 
that the adult defi brillation pads be placed in the anterior-anterior 
(front-front) position. 

WARNINGS

 (continued)

•  Some very low amplitude or low frequency VF rhythms may not 

be interpreted as shockable. Some very low amplitude or low 
frequency VT rhythms may not be interpreted as shockable.

•  Handling or transporting the patient in any way during ECG analysis 

can cause incorrect or delayed diagnosis, especially if very low 
amplitude or low frequency rhythms are present.  If the patient is 
being transported, stop vehicle before beginning ECG analysis.

•  In patients with cardiac pacemakers, the DDU-2000 Series 

AED may have reduced sensitivity and not detect all shockable 
rhythms. If you know the patient has an implanted pacemaker, do 
not place electrodes directly over an implanted device. 

•  During defi brillation, air pockets between the skin and 

defi brillation pads can cause patient skin burns. To help 
prevent air pockets, make sure self-adhesive defi brillation pads 
completely adhere to the skin. Do not use dried out or expired 
defi brillation pads.

•  Defi brillation may cause skin burns around the defi brillation pads 

area.

•  User-initiated and automatic self-tests are designed to assess the 

DDU-2000 Series AED’s readiness for use. However, no degree 
of testing can assure performance or detect abuse, damage, or a 
defect that occurred after the most recent test is completed.

•  Use of damaged equipment or accessories may cause the device to 

perform improperly and/or result in injury to the patient or operator.

•  Possible misinterpretation of ECG data.  The frequency response 

of the LCD display is intended for basic ECG rhythm identifi cation; 
it does not provide the resolution required for pacemaker pulse 
identifi cation or accurate measurements, such as QRS duration 
and ST segment interpretation.  For such purposes an ECG 
Monitor with an appropriate frequency response should be used. 

•  Follow voice prompts if the LCD screen becomes blank or 

unreadable.

•  It may be possible for the AED to not detect a shockable rhythm, 

not deliver a shock to a shockable rhythm or not deliver the 
intended energy during defi brillation.

•  It may be possible that the AED recommends a shock for a 

non-shockable rhythm, and if a shock is delivered, VF or cardiac 
arrest may occur.

•  Even if defi brillation occurs, the sudden cardiac arrest event may 

not result in survival.

•  Defi brillation may cause myocardial damage or post-shock 

dysfunction.

•  Therapy cannot be delivered while an AED software update is in 

process.

•  Do not turn off the AED or remove the battery pack or the update 

data card until an AED software update process is complete 
as these actions may render the AED incapable of delivering 
therapy. If any of these interruptions occur, restart the update 
procedure from the beginning.

WARNINGS

 (continued)

•  While in demonstration mode, the AED cannot perform a rescue. 

If at any time during the demonstration there is a need to perform 
a rescue, press and hold the green on/off button for two seconds 
to power off the AED. Then press the green on/off button to power 
on the unit and begin a rescue protocol. The demonstration card 
does not need to be removed in order to perform a rescue.

!

CAUTIONS:

Conditions, hazards, or unsafe practices that 

may result in minor personal injury, damage to 

the DDU-2000 Series AED, or loss of data.

•  Follow all battery pack labeling instructions. Do not install battery 

packs after the expiration date.

•  Follow all defi brillation pad label instructions. Use defi brillation 

pads prior to their expiration date.

•  The defi brillation pads should not be in continuous contact with 

the patient’s skin for more than 24 hours.

•  Allergic dermatitis or a minor skin rash may result in patients that are 

sensitive to the materials used for the defi brillation pads.  Remove 
the defi brillation pads from the patient as soon as practical.

•  Recycle or dispose of lithium battery packs in accordance with 

local, state, provincial, and/or national regulations. 
To avoid fi re and explosion hazard, do not burn or incinerate the 
battery pack.  Do not crush.  

•  Use and store the DDU-2000 Series AED only within the range of 

environmental conditions specifi ed in the technical specifi cations.

•  If possible, disconnect the DDU-2000 Series AED from the patient 

prior to use of other defi brillators.

•  Using non-Defi btech Data Cards (DDC cards) may damage the 

unit and will void the warranty. 

•  Defi bView software is not intended for clinical use. Information 

presented by Defi bView should not be used for making clinical 
decisions.

•  Although the DDU-2000 Series AED is designed for a wide variety 

of fi eld use conditions, rough handling beyond specifi cations may 
result in damage to the unit.

•  Federal Law (USA) restricts this device to sale by or on the order 

of a physician.

Refer to User Manual Section 9.1 for 

Summary of Primary Clinical Studies 

and Potential Adverse Effects 

of the Device on Health.

Summary of Contents for DDU-2200

Page 1: ...ance source for complete technical specifications Defibtech Automated External Defibrillator Lifeline ReviveR VIEW AUTO DDU 2200 Lifeline ReviveR VIEW DDU 2300 Lifeline ReviveR ECG DDU 2450 Lifeline R...

Page 2: ...1 2 3 PRESS ON BUTTON APPLY PADS FOLLOW AED INSTRUCTIONS 2 DAC U2510EN BF rev H...

Page 3: ...and data used in any examples are fictitious unless otherwise noted Limited Warranty The Limited Warranty shipped with Defibtech AED products serves as the sole and exclusive warranty provided by Defi...

Page 4: ...Mode 18 4 1 Overview 18 4 2 Preparation 20 4 3 Heart Rhythm Analysis 23 4 4 Delivering The Shock 23 4 5 CPR Period 23 4 6 Post Use Procedures 24 4 7 AED Mode Voice And Text Prompts 24 4 8 Operational...

Page 5: ...loading The Internal Data Log 49 9 Technical Specifications 50 9 1 Defibtech DDU 2000 Series AED 50 9 2 Battery Packs 59 9 3 Self Adhesive Defibrillation Pads 59 9 4 Event Documentation 59 9 5 Defibte...

Page 6: ...ates in AED Mode DDU 2450 Operates in AED Mode includes patient ECG display DDU 2475 Equivalent to the DDU 2450 configurable to include additional features contact Defibtech or your distributor for de...

Page 7: ...though the duration of the generated waveform will vary Defibrillation and AED operating power is supplied with a replaceable non rechargeable battery pack that provides for long standby life and low...

Page 8: ...unit needs attention from the user or needs servicing G Softkey Buttons Three context sensitive softkey buttons are used to navigate menus or select actions H USB Port The USB port is provided to per...

Page 9: ...OFF Button C Display Screen B SHOCK Button Semi Automatic units ONLY SHOCK Required Indicator on DDU 2200 A Speaker F Active Status Indicator ASI G Softkey Buttons Top Center Bottom H USB Port I Defi...

Page 10: ...pads model number DDP 2002 if available Contraindications Lifeline ReviveR VIEW AUTO DDU 2200 Lifeline ReviveR VIEW DDU 2300 Lifeline ReviveR ECG DDU 2450 Lifeline ReviveR ECG DDU 2475 Automated Exter...

Page 11: ...o modification of this equipment is allowed Electrical Shock Hazard Dangerous high voltages and currents are present Do not open unit remove cover or back or attempt repair There are no user serviceab...

Page 12: ...CG analysis can cause incorrect or delayed diagnosis especially if very low amplitude or low frequency rhythms are present If the patient is being transported stop vehicle before beginning ECG analysi...

Page 13: ...ration date The defibrillation pads should not be in continuous contact with the patient s skin for more than 24 hours Allergic dermatitis or a minor skin rash may result in patients that are sensitiv...

Page 14: ...to begin using the device 3 1 Overview The following components and accessories are included with your DDU 2000 Series AED Replacement and other accessories are detailed in the DDU 2000 Series AED Acc...

Page 15: ...tential cross infection improper performance of the device inadequate delivery of therapy and or injury to the patient or operator 3 3 Installing And Removing The Battery Pack The battery pack provide...

Page 16: ...ion will be off In this case immediately replace the battery pack or reinsert it into the unit to restore active status indication Active Status Indicator ASI Flashing Green The DDU 2000 Series AED is...

Page 17: ...latch engages Note DDU 2000 Series AEDs running software version 1 3 or below may not function with certain DDC Cards Contact Defibtech for support refer to Chapter 12 for contact information CAUTION...

Page 18: ...Series In AED Mode This chapter describes how to use the DDU 2000 Series in AED Mode In AED Mode the unit analyzes the patient s rhythm and automatically charges if a shockable rhythm is detected Push...

Page 19: ...y button to select CPR coaching with compressions only no breathing or CPR coaching with compressions and breathing Information Softkey Icon When this icon is present on the screen the user may press...

Page 20: ...etely discharged or is not installed in the unit the active status indication will not be available In this case immediately replace the battery pack or reinsert it into the unit to restore active sta...

Page 21: ...hile the pads themselves remain sealed in their package This reduces the time needed to set up and start treatment in an emergency The Defibtech AED should be stored with the pads connector attached t...

Page 22: ...r 8 years or less than 55 pounds 25 kg is different than placement for adults and children 8 years or older or over 55 pounds 25 kg Do not delay therapy to determine exact age or weight If pediatric p...

Page 23: ...shock Also if the SHOCK button is not pressed within 30 seconds of the initial Press Flashing Shock Button prompt the unit will automatically cancel the shock DDU 2200 AED If the fully automatic DDU...

Page 24: ...ge of 8 or less than 55 pounds 25 kg For children 8 years or older or over 55 pounds 25 kg and for adults adult pads should be used Do not delay therapy to determine exact age or weight Training pads...

Page 25: ...that the pads are not touching each other and that the patient is dry If the prompts continue try replacing the pads with a new set Pausing for CPR Pausing for CPR Purpose If too long a period of time...

Page 26: ...cts a rhythm change to a non shockable one charging will abort and the user will be prompted to begin CPR Note At any time during the charging process or after the AED has been charged the operator ma...

Page 27: ...en The factory default setting provides prompts for chest compressions only no breathing Breathing prompts can be included either by changing the menu option refer to the CPR Breathing section in Chap...

Page 28: ...ient for rescue breaths and to give breaths Breathe Breathe Purpose This instructs the user to give rescue breaths Each time the instruction is given the user should give the patient a rescue breath 4...

Page 29: ...ll the most likely cause is that the battery pack needs to be replaced Refer to the Installing And Removing The Battery Pack section in Chapter 3 of this manual Once the battery pack has been replaced...

Page 30: ...test failed Service code xxxx Power OnTest Failed Service Code xxxx Purpose This indicates that the DDU 2000 Series AED has failed the power on self test and may be non operational and may require se...

Page 31: ...check to ensure that the pads package has not expired Do not use pads past their expiration date Discard expired pads Next check to ensure that the pads package has not been torn opened or damaged Di...

Page 32: ...d to the patient s care provider A new DDC card should be installed before the next use To remove the DDC card ensure the AED is OFF Locate the data card USB port access door on the right hand side of...

Page 33: ...ines to be adhered to when cleaning the device The battery pack should be installed when cleaning the DDU 2000 Series AED Do not immerse the DDU 2000 Series AED in fluids or allow fluids to enter the...

Page 34: ...ries AED Serial Number ___________________________________________________ Defibtech DDU 2000 Series AED Location __________________________________________________________ Date Check unit and accesso...

Page 35: ...SI does not flash at all while unit is in standby powered off Battery pack not inserted Insert battery pack Battery pack is low or needs servicing Replace battery pack or call for service Unit needs s...

Page 36: ...op motion prompt Patient motion has been detected Stop patient motion Stop interference prompt External interference has been detected Stop external interference Analyzing interrupted prompt Motion or...

Page 37: ...tons Top Center Bottom Rescue now AED status AED maintenance AED options Rescue options Help topics Turn AED off AED Main Menu Title The name of the menu appears in a banner across the top of the scre...

Page 38: ...D options Rescue options Help topics Turn AED off AED Main Menu Rescue now Puts device in AED Mode AED status Displays current AED status information AED maintenance Displays AED Maintenance Menu scre...

Page 39: ...it is in Maintenance Mode To enter Navigate to Perform AED test AED Main Menu AED maintenance Perform AED test What it does When the user selects Perform AED test selection and presses the CENTER soft...

Page 40: ...btech Data Card DDC card To exit When the unit finishes performing the AED upgrade follow the displayed and spoken instructions Transfer DataTo Card Transfer data to card will initiate a data transfer...

Page 41: ...it will speak and display Data Card Missing Refer to the InstallingThe Defibtech Data Card DDC card section in Chapter 3 of this manual To exit When the unit finishes formatting the data card it will...

Page 42: ...roll to and highlight the selection Go to main menu Press the CENTER softkey button The unit will exit the AED options menu screen and return to the AED Main Menu SystemTime The System time option all...

Page 43: ...ance Mode To enter Navigate to Volume level AED Main Menu AED options Volume level What it does The Volume level option allows the user to set AED audio to high medium or low volume Once the Volume le...

Page 44: ...eathing select the Rescue options from the AED Main Menu screen Before you begin Be sure the unit is in Maintenance Mode To enter Navigate to Rescue options AED Main Menu Rescue options What it does T...

Page 45: ...l protocol code This code is a special code that encodes all the important information regarding the protocol i e number of compressions and breaths per cycle compression pacing number of CPR cycles a...

Page 46: ...ent Analyzing and shocking Performing CPR Replacing the battery Replacing the pads Checking the AED status Replacing the data card Use the TOP and BOTTOM softkey buttons to scroll through the differen...

Page 47: ...more information please visit www defibtech com or contact Defibtech or your distributor 7 4 Data Cards The DDU 2000 Series AED is designed to optionally use Defibtech Data Cards DDC cards The AED wi...

Page 48: ...intenance software The AED has a mini USB connector located on the right side of the unit behind the data card USB port access door WARNING Do not connect the DDU 2000 Series AED to a PC or other devi...

Page 49: ...delivery information This information can be reviewed using the DefibView application CAUTION Using non Defibtech Data Cards DDC cards may damage the unit and will void the warranty 8 3 Downloading T...

Page 50: ...he end of battery life and for temperatures below 10 C Charge time from the initiation of rhythm analysis to readiness for discharge Meets or exceeds AAMI DF80 and IEC 60601 2 4 requirements Charge ti...

Page 51: ...se to special arrhythmias Ventricular bigeminy A1 80 bpm Slow alternating ventricular bigeminy A2 30 bpm Rapid alternating ventricular bigeminy A3 120 bpm Bidirectional systoles A4 90 bpm Pacemaker pu...

Page 52: ...5 2 8 ms 2 8 ms 50 4 1 ms 4 1 ms 75 7 2 ms 4 8 ms 100 9 0 ms 6 0 ms 125 12 0 ms 8 0 ms 150 12 0 ms 8 0 ms 175 12 0 ms 8 0 ms Energy AED Mode DDP 2001 Adult Defibrillation Pads Nominal Nominal Energy L...

Page 53: ...ions Category M From room temperature to temperature extreme one hour duration updated specification for DDU 2000 Series AEDs running software revision 2 4 or above Patient Analysis System The DDU 200...

Page 54: ...LC submitted a comparison of the Defibtech adult and pediatric defibrillation waveforms for the DDU 100 and DDU 2000 series AEDs and the Philips defibrillation waveforms that were also used for the or...

Page 55: ...or type of responder A summary of the results is presented in the following table Biphasic vs Monophasic Waveform Biphasic Patients Number Monophasic Patients Number P Value Defibrillation Efficacy 1...

Page 56: ...eous circulation without post shock dysfunction in this pediatric model Atkins et al is a post market observational study on pediatric patients intended to evaluate reported uses of pediatric pads tha...

Page 57: ...s or exceeds IEC 60601 2 4 requirements meets the AAMI DF80 requirement and the AHA recommendation2 of Sensitivity 75 Non Shockable Rhythm Normal Sinus Rhythm 213 100 100 Meets or exceeds IEC 60601 2...

Page 58: ...ummary 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Circulation 2015 132 suppl 2 S315 S367 10 Monsieurs KJ et al on behalf of t...

Page 59: ...inches2 50 cm2 each nominal Cable connector type Integrated Integrated Cable length 4 feet 122 cm typical 4 feet 122 cm typical Expiration date 2 5 years from date of manufacture 2 5 years from date...

Page 60: ...only be disposed of in compliance with the provisions of the above mentioned European directive and as amended as well as with the corresponding national regulations Severe penalties are possible for...

Page 61: ...ons IEC 61000 3 2 Not applicable Battery operated equipment Voltage fluctuations flicker emissions IEC 61000 3 3 Not applicable Battery operated equipment ELECTROMAGNETIC IMMUNITY Immunity test IEC 60...

Page 62: ...ssess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the DDU 2000 Series AED i...

Page 63: ...This device may not cause harmful interference and 2 This device must accept any interference received including interference that may cause undesired operation This equipment has been tested and fou...

Page 64: ...Y Y Y Y Consult operating instructions Refer to instruction manual booklet X X X X Y Y Y Y Do not damage or crush X X X X Y Y Y Y Follow proper disposal procedures Meets the requirements of the Europ...

Page 65: ...tricts this device to sale by or on the order of a physician Y Y Y Y Catalogue number Keep dry Handle with care Transportation and storage requirements See environmental requirements on packaging LATE...

Page 66: ...g service defibtech com Service and Repair This product and its accessories are manufactured and sold under license to at least one or more of the following United States patents 5 591 213 5 593 427 5...

Page 67: ...External Defibrillator Lifeline ReviveR VIEW AUTO DDU 2200 Lifeline ReviveR VIEW DDU 2300 Lifeline ReviveR ECG DDU 2450 Lifeline ReviveR ECG DDU 2475 Operating Guide For concise guidance on set up use...

Page 68: ...ies Notices Defibtech L L C shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material Information...

Page 69: ...n 55 lbs 25 kg use Defibtech child infant defibrillation pads model number DDP 2002 if available CONTRAINDICATIONS Lifeline ReviveR VIEW AUTO DDU 2200 Lifeline ReviveR VIEW DDU 2300 Lifeline ReviveR E...

Page 70: ...t Display Screen Displays video text prompts and information Speaker Defibtech Data Card DDC Card Optional USB Port DEFIBRILLATION PADS PACKAGE BATTERY PACK BACK OF AED FRONT OF AED Defibtech Automate...

Page 71: ...tery pack test Wait for the test to complete and for the unit to turn off For more information refer to page 16 3 4 CHECKTHE STATUS READTHE USER MANUAL at www defibtech com For more information refer...

Page 72: ...hest and remove excess chest hair 3 4 PREPARE PADS PLACE PADS Peel adhesive pads from blue liner Tear open pads package Apply pads to patient s bare chest For more information refer to page 15 ADULT P...

Page 73: ...button has been pressed the fully automatic DDU 2200 AED will shock the patient automatically without any user intervention Do not touch the patient while the SHOCK Required auto indicator is flashin...

Page 74: ...l it is fully seated in the unit The connector will only fit in one way if the connector does not fit rotate the connector before trying again The connected pads package should be stored in the pad st...

Page 75: ...w may not function with certain DDC Cards Contact Defibtech for support refer to page 32 for contact information CAUTION Using non Defibtech Data Cards DDC cards may damage the unit and will void the...

Page 76: ...ed actions such as viewing unit information initiating unit self tests changing unit parameters downloading rescue data and upgrading software To enter Maintenance Mode press the bottom softkey button...

Page 77: ...sfer data to card Format data card Run application from card Go to main menu AED Maintenence Rescue now Preparing the patient Analyzing and shocking Performing CPR Replacing the battery Replacing the...

Page 78: ...place battery pack as soon as possible Display screen does not work Battery pack depleted Replace battery pack Battery pack not inserted properly Make sure battery pack is oriented correctly and fully...

Page 79: ...r delayed diagnosis by the patient analysis system Do not place adult defibrillation pads in the anterior posterior front back position A shock or no shock decision may be inappropriately advised The...

Page 80: ...Adult 12 inches2 77cm2 nominal each pad Child Infant 7 75 inches2 50cm2 nominal each pad TYPE Pre connected single use non polarized disposable self adhesive electrodes with cable and connector DEFIB...

Page 81: ...uirements of the European Medical Device Directive Meets the requirements of the Radio Equipment andTelecommunications Directive 1999 5 EC Classified byTUV Rheinland of NA with respect to electric sho...

Page 82: ...ncluding persons not authorized by Defibtech LLC specified maintenance is not performed the defibrillator is used with one or more unauthorized accessories the associated accessories are used with an...

Page 83: ...ice Reporting service defibtech com Service and Repair This product and its accessories are manufactured and sold under license to at least one or more of the following United States patents 5 591 213...

Reviews: