Rescue Protocol
The AED supports two rescue protocols at one time. The
Rescue protocol
option allows the user to
select a rescue protocol. Rescue protocol options include the AHA/ERC protocol or “Custom”
Before you begin
: Be sure the unit is in Maintenance Mode.
To enter
: Navigate to
Rescue protocol
:
AED Main Menu
Ú
Rescue options
Ú
Rescue protocol
What it does
: The
Rescue protocol
option allows the user to select between up to two rescue protocols
that have been enabled in the unit. The factory default setting of the rescue protocol is “
AHA
.”
To change the protocol, press the CENTER softkey button to highlight the protocol. The user will be
prompted to enter a password to proceed. The password may be obtained from your medical director
or from Defibtech. (For Defibtech contact information, please refer to the
“Contacts”
section in
Chapter 12.) Once the password has been entered, the user may select between the two protocols.
To enter the password use the TOP softkey button to scroll through numbers. Once the correct number
appears, use the CENTER softkey button to move to the next space. Once all of the numbers have been
entered press the CENTER softkey button. The user will now be able to choose a different rescue protocol.
Settings
Before you begin
: Be sure the unit is in Maintenance Mode.
To enter
: Navigate to
Settings
:
AED Main Menu
Ú
Rescue options
Ú
Settings
What it does
: The
Settings
option allows the user to change the currently enabled
protocol by entering a special protocol code. This code is a special code that encodes all
the important information regarding the protocol (i.e. number of compressions and breaths
per cycle, compression pacing, number of CPR cycles, and number of shocks between
CPR). This code is custom generated by Defibtech. If the code is not entered correctly the
protocol will not be changed. Based on the protocol code entered, the currently selected
protocol will be changed to that described by the special protocol code. You can obtain this
code from your medical director or from Defibtech. (For Defibtech contact information,
please refer to the “
Contacts
” section in Chapter 12.) Once the code has been entered, the
settings will have been changed.
To enter the code use the TOP softkey button to scroll through numbers/letters. Once
the correct number/letter appears, use the CENTER softkey button to move to the next
space. Once all of the numbers/letters have been entered, press the CENTER softkey
button. The settings will have been changed based on the code entered.
Default View
(DDU-2450 only)
The
Default view
option allows the user to select Video or ECG as the default view when the AED
is powered.
Before you begin
: Be sure the unit is in Maintenance Mode.
To enter
: Navigate to
Default view
:
AED Main Menu
Ú
Rescue options
Ú
Default view
What it does
: Selects which view the AED presents on power up.
Use the TOP and BOTTOM softkey buttons to select the desired view. When the desired selection is
chosen, press the CENTER softkey button to set the feature. The AED will now use that default view
setting.
Note
: The DDU-2450's factory default setting of Default View is “
Video
.”