For more detailed information, refer to the User Manual (at www.defi btech.com).
USING THE AED - ECG DISPLAY
(DDU-2450 ONLY)
SELECTING ECG DISPLAY
The DDU-2450 allows the user to display
the patient ECG when the unit is being
used as an AED.
To select ECG display, press the
bottom softkey button next to
the
Mode Select Icon
(shown
at left) to bring up the
Mode of
Operation
screen (shown at right).
Press the corresponding softkey (bottom
button) to select
AED / ECG Mode
.
Mode of Operation screen
(with corresponding softkey buttons)
Mode of Operation
AED / Video Mode
AED / ECG Mode
ECG Display screen
Analyzing
!
WARNING
Possible misinterpretation of ECG data. The frequency
response of the LCD display is intended for basic ECG
rhythm identifi cation; it does not provide the resolution
required for pacemaker pulse identifi cation or accurate
measurements, such as QRS duration and ST segment
interpretation. For such purposes an ECG Monitor with an
appropriate frequency response should be used.
THE DEFIBRILLATION PADS
HOW TO CONNECT THE PADS
Insert the connector end of the defi brillation pad cable into the pads
connector socket on the top-left corner of the DDU-2000 Series AED as
shown. Insert pads connector fi rmly until it is fully seated in the unit. The
connector will only fi t in one way – if the connector does not fi t, rotate the
connector before trying again.
The connected pads package should be stored in the pad storage area
on the back of the DDU-2000 Series AED (see diagram on page 7).
After connecting the pads connector to the unit, push the pads pack-
age, rounded end fi rst, with the pictures on the package facing out, into the pad storage area.
When the pads package is fully inserted, press the pad cable into the groove in the back of the
unit to hold the cable in place and tuck any excess cable behind the pads package.
WHICH PADS TO USE
WHEN TO REPLACE THE PADS
The Defi btech defi brillation pads are intended for one time use only. The pads must be
replaced after each use or if the package has been damaged.
It is important to check the expiration date of the pads.
Do not use pads past their expiration
date
. Discard expired pads.
Use only Defi btech defi brillation pads
.
You may check the status of the pads when the unit is off by pressing the center softkey
button to display the AED Status screen.
The expiration date is also printed on the outside of the sealed package.
For adults and children
8 years or older or over
55 pounds (25 kg),
use adult pads
For infants and children under 8 years
or less than 55 pounds (25 kg), use child/
infant pads (Note: child/infant pads can be
identifi ed by their blue connector
and blue pads package)
ADULT PADS
CHILD/INFANT PADS