DAC-U2530EN-AG rev E
DAC-U2530EN-AG rev E
16
17
For more detailed information, refer to the User Manual (at www.defi btech.com).
THE BATTERY PACK
THE DEFIBTECH DATA CARD
(OPTIONAL)
HOW TO INSERT AND REMOVE THE DEFIBTECH DATA CARD
The optional Defi btech Data Card (DDC card) is used to store event and audio information
collected by the AED. All DDU-2000 Series AEDs will operate without DDC cards and will still
store select event information internally. Information stored on the DDC card is retrievable
with a separate Defi btech PC-based software package. (Refer to the
“Defi bView”
section of
the DDU-2000 Series User Manual available at www.defi btech.com.)
Before installing the DDC card, ensure the AED
is turned OFF. Locate the data card/USB port
access door on the right-hand side of the unit.
Open the data card/USB port access door by
slightly pushing and then sliding the door down
to release the latch. The door will spring open.
Insert the DDC card into the thin slot in the
side of the AED centered above the USB port
opening, notched end fi rst, label side up, until it
clicks into place. The card should be fl ush with
the surface of the slot. If the card does not push
in all the way, it may have been inserted upside
down. In that case, remove the card, fl ip it over,
and try inserting it again.
To remove the DDC card, press the card as far as it will go and then release. Upon release,
the DDC card will be partially ejected and can be removed by pulling the DDC card the rest
of the way out.
Close the data card/USB port access door by closing and then pushing the door up until the
door latch engages.
Note:
DDU-2000 Series AEDs running software version 1.3 or below may not function with
certain DDC Cards. Contact Defi btech for support (refer to page 32 for contact information).
!
CAUTION
Using non-Defi btech Data Cards (DDC cards) may damage
the unit and will void the warranty.
XX
XX
XX
RA
LA
LL
HOW TO INSERT AND
REMOVE THE BATTERY PACK
Before inserting the battery pack into the DDU-
2000 Series AED, ensure that the battery pack
opening in the back of the AED is clean and clear
of any foreign objects. Insert the battery pack
into the opening in the back of the AED.
Push the pack all the way in until the latch clicks.
The battery pack will only fi t in one way. If the
battery pack does not fi t, rotate the battery pack
before trying again. Once fully inserted, the
battery pack surface should be fl ush with the
back of the AED. Within moments of insertion,
the DDU-2000 Series AED will turn on and run
a battery pack insertion test. When the test is
completed, the unit will report the status of the
battery pack and shut down. (
Note:
The battery
pack must have been removed from the unit for
at least 10 seconds for the battery pack self-test
to be performed automatically.)
To remove the battery pack, push the battery
pack eject release latch. After the battery pack
is partially ejected, pull the battery pack out.
WHEN TO REPLACE THE BATTERY PACK
It is important to check the expiration date of the
battery pack.
The battery pack should be used
before the expiration date
. When the battery
pack is low, the unit will indicate “battery low”
or “replace battery now” and the Active Status
Indicator will fl ash red. The battery pack should
be replaced immediately.
Use only Defi btech
battery packs
.
You may check the status of the battery pack
when the unit is off by pressing the center
softkey button to display the AED Status Screen.
The expiration date is also printed on the label on
the battery pack as shown at right.
Battery Pack Label
Located Here