If running other applications:
WARNING
While in demonstration mode, the AED cannot perform a rescue. If at any time
during the demonstration there is a need to perform a rescue, press and hold
the green on/off button for two seconds to power off the AED. Then press
the green on/off button to power on the unit and begin a rescue protocol. The
demonstration card does not need to be removed in order to perform a rescue.
Note
:
If a data card is not inserted, the unit will speak and display “Data Card Missing.”
(Refer to the “
Installing The Defibtech Data Card (DDC card)
” section in Chapter 3 of this manual.)
To exit
: When the unit finishes performing the application, follow the displayed and spoken instructions.
6.7 AED Options Screen
To manually configure AED options such as time, date, volume, and audio recording, select the
AED options
from the AED Main Menu screen.
Before you begin
: Be sure the unit is in Maintenance Mode.
To enter
: Navigate to
AED options
:
AED Main Menu
Ú
AED options
What it does
: The unit will display the AED options menu screen. This screen will allow the user to modify the
following user changeable parameters:
• System time
• System date
• Volume level
• Audio recording
• Status broadcast
To exit:
Using the TOP or BOTTOM softkey buttons, scroll to and highlight the selection.
Go to main menu
. Press the CENTER softkey button. The unit will exit the AED options menu screen and
return to the AED Main Menu.
System Time
The
System time
option allows the user to set the time of the internal AED clock.
Before you begin
: Be sure the unit is in Maintenance Mode.
To enter
: Navigate to
System time
:
AED Main Menu
Ú
AED options
Ú
System time
What it does
: The
System time
option allows the user to set the time of internal AED clock (using
the 24 hour clock). Once the
System time
option is selected, press the CENTER softkey button to
enter set time mode:
The hours selection will be highlighted green:
• Press the TOP or BOTTOM softkey buttons to adjust the hours to the desired time.
• Press the CENTER softkey button to accept the hours setting.
The minutes selection will be highlighted green:
• Press the TOP or BOTTOM softkey buttons to adjust the minutes to the desired time.
• Press the CENTER softkey button to accept the minutes setting.
The seconds selection will be highlighted green:
• Press the TOP or BOTTOM softkey buttons to adjust the seconds to the desired time.
• Press the CENTER softkey button to accept the seconds setting.
The time is now set and the user may use TOP or BOTTOM softkey buttons to navigate
to other choices in the menu.
Note
: The factory default setting of the internal AED clock is Universal Time (GMT).
6. Maintenance Mode
Run Application From Card (continued)
42
DAC-U2510EN-BF rev H