DAC-U2530EN-AG rev E
DAC-U2530EN-AG rev E
8
9
For more detailed information, refer to the User Manual (at www.defi btech.com).
SETTING UP THE AED
The DDU-2000 Series AED is designed to be stored in a “ready”
state so that few steps are required to begin using the AED.
1
2
CONNECT
THE PADS TO
THE AED
INSTALL THE
BATTERY PACK
Ensure that the pads package has not expired.
Expired pads must not be used.
For more information, refer to page 15.
Ensure that the battery pack has not expired.
Expired battery packs must not be used.
When the battery pack is installed, the AED will turn
on and run a battery pack test. Wait for the test to
complete and for the unit to turn off.
For more information, refer to page 16.
3
4
CHECK THE
STATUS
READ THE
USER MANUAL
(at www.defi btech.com)
For more information, refer to page 18.
With the AED off
, press
and release the CENTER
softkey button. Ensure
that the AED Status
screen appears and that
the “AED Status” is “OK”.
AND
When the AED is off
, the Active
Status Indicator (ASI) should fl ash
green. If the ASI fl ashes red, is
solid red, or if there is no fl ashing
light, the unit requires service.
Comprehensive information about the
DDU-2000 Series AED can be found in the
User Manual (at www.defi btech.com).
AED Status screen
AED status
OK
Battery status
Expires 01/20xx
Pads status
Adult
Expires 06/20xx
AED S/N
099999999
Battery S/N
099999999
Software version
Vx.x