• Therapy cannot be delivered while an AED software update is in process.
• Do not turn off the AED or remove the battery pack or the update data card until an AED software update process is
complete as these actions may render the AED incapable of delivering therapy. If any of these interruptions occur,
restart the update procedure from the beginning.
• While in demonstration mode, the AED cannot perform a rescue. If at any time during the demonstration there is a
need to perform a rescue, press and hold the green on/off button for two seconds to power off the AED. Then press
the green on/off button to power on the unit and begin a rescue protocol. The demonstration card does not need to
be removed in order to perform a rescue.
2.2
CAUTION
!
DANGER
!
WARNING
!
CAUTIONS
:
Conditions, hazards, or unsafe practices that may result in minor
personal injury, damage to the DDU-2000 Series AED, or loss of data.
• Follow all battery pack labeling instructions. Do not install battery packs after the expiration date.
• Follow all defibrillation pad label instructions. Use defibrillation pads prior to their expiration date.
• The defibrillation pads should not be in continuous contact with the patient’s skin for more than 24 hours.
• Allergic dermatitis or a minor skin rash may result in patients that are sensitive to the materials used for the
defibrillation pads. Remove the defibrillation pads from the patient as soon as practical.
• Recycle or dispose of lithium battery packs in accordance with local, state, provincial, and/or national regulations. To
avoid fire and explosion hazard, do not burn or incinerate the battery pack. Do not crush.
• Use and store the DDU-2000 Series AED only within the range of environmental conditions specified in the technical
specifications.
• If possible, disconnect the DDU-2000 Series AED from the patient prior to use of other defibrillators.
• Using non-Defibtech Data Cards (DDC cards) may damage the unit and will void the warranty.
• DefibView software is not intended for clinical use. Information presented by DefibView should not be used for
making clinical decisions.
• Although the DDU-2000 Series AED is designed for a wide variety of field use conditions, rough handling
beyond specifications may result in damage to the unit.
• Federal Law (USA) restricts this device to sale by or on the order of a physician.
WARNINGS (continued)