ADD-ONS
67
5
CAUTION
Channel balancing is performed at the factory and should never be
changed during the camera life.
This procedure should be performed
only in emergency situations and by qualified personnel.
To perform this procedure click on the “Channel Balancing” icon on the Genius™ toolbar and
follow the steps given.
CAUTION
Allow the camera to warm-up for at least 15 minutes before performing
this procedure. If the storage temperature of the camera is different from
the working one, this time has to be increased.
Place the Channel Balancing Chart (93ACC1380) in front of the camera at the distance with
the lowest illumination (usually maximum distance).
Click the "Next" button to continue. At this point the Oscilloscope window appears.
Make sure you see that the target in the oscilloscope window is like Figure 50.
Figure 50 – Oscilloscope View: Gain/Offset Target