ADD-ONS
69
5
Figure 51 – Interpixel Equalization: Black Cover
Click the "Exit" icon in the upper right of the oscilloscope window and wait for the procedure
to automatically sample the image.
As prompted, remove the object and place the white reference target (93ACC1370) at the
indicated distance.
CAUTION
The reference chart should be perfectly clean and flat. Any dust, fold or
crease on this target will alter the results of the procedure.
Clean the target with a soft cloth and alcohol or substitute it.
High-resolution cameras are the most sensitive. While for low-resolution cameras any
uniform white paper/plastic stripe can be used as a target, for these kind of cameras the
standard reference target is recommended.
Click the "Next" button to continue. The Oscilloscope window reappears. Verify that the
image is as uniform as possible.
As prompted, move the white target a few millimeters and repeat the acquisition. The
procedure will be repeated a few times as prompted by the software.
During the measurement period be careful not to move the camera or change the
illumination.