Dakota Ultrasonics
88
If we consider the initial probe zero value constant, as the hardware cal bars or glass
block have not been replaced, the future verifications should always show an
elongation value of +/- 0.0005”, in order to be considered “In Spec”. If the probe
zero/cal is out of spec, the user should redo the zero/cal procedure in its entirety.
The next section outlines the verification procedures.
Verifying the Probe Zero at a Later Time
1) Press the
key once to activate the menu items tab. Press the
key
multiple times to tab right, and the
key multiple times to tab left, until the
DATA
menu is highlighted and displaying the submenu items.
2) Press the
and
arrow keys multiple times to scroll through the sub
menu items until
OPEN
is highlighted.
3) Press the
key to display the
Open Group List Box.