![DAKOTA ULTRASONICS MiniMax v2.0 Operation Manual Download Page 77](http://html1.mh-extra.com/html/dakota-ultrasonics/minimax-v2-0/minimax-v2-0_operation-manual_3268519077.webp)
MiniMax
Bolt Tension Monitor
71
22) Repeat steps
20 & 21
until the
TEMP COEF
value is correctly displayed.
23) Press the
key to return to the
One Point Zero List
, or
to cancel
entering the
TEMP COEF
value.
24) Press the
and
arrow keys multiple times to scroll through the
One
Point Zero List Items
until
ULTRASONIC 1
is highlighted.
25) Press the
key to display the measurement screen.
26) Press the
key to automatically adjust the settings and measure the
calibration standard.
27) Press the
and
arrow keys multiple times to scroll the hot menu
items until
ACCEPT CAL
is highlighted.
28) Press the
key to accept the calibration and return to the
One Point
Zero
list, or
to cancel the calibration.
29) Press the
and
arrow keys multiple times to scroll through the
One
Point Zero
list until
CALC ZERO
is highlighted.