16) Repeat steps
14 & 15
until the
PHY LEN 2
value is correctly displayed.
17) Press the
key to return to the
Two Point Zero
list, or
to cancel
entering the
PHY LEN 2
.
18) Press the
and
arrow keys to scroll through the
Two Point Zero
list until
VELOCITY
is highlighted.
Note:
This refers the velocity of the probe zero cal bars. Since there is a
velocity for each cal bar, use the velocity of the longest cal bar as it will be most
significant – 6” bar. If the glass block is being used, it will have only one
velocity specified for the block.
19) Press the
key to display the
Digits Edit Box
.
20) Press the
and
arrow keys multiple times to scroll the highlighted
value.
21) Press the
and
arrow keys multiple times to scroll the digit
locations.
22) Repeat steps
20 & 21
until the
VELOCITY
value is correctly displayed.
23) Press the
key to return to the
Two Point Zero
list, or
to cancel
entering the
VELOCITY
.