![DAKOTA ULTRASONICS MiniMax v2.0 Operation Manual Download Page 61](http://html1.mh-extra.com/html/dakota-ultrasonics/minimax-v2-0/minimax-v2-0_operation-manual_3268519061.webp)
MiniMax
Bolt Tension Monitor
55
1) Press the
key multiple times to tab right and the
key multiple times
to tab left until the
CAL
menu is highlighted and displaying the submenu
items.
2) Press the
and
arrow keys multiple times to scroll through the sub
menu items until
ZERO MODE
is highlighted.
3) Press the
and
arrow keys multiple times to scroll the zero mode
options until the
AUTO
option is displayed.
Note:
The auto option will perform a two-point zero calibration using the first
and second multiple of the calibration standard. Therefore, if a 3” bar is used,
the second point will be at 6”.
4) Press the
and
arrow keys multiple times to scroll through the sub
menu items until
MEASURE ZERO
is highlighted.