MiniMax
Bolt Tension Monitor
75
procedures in the next section demonstrate a
Two-Point
calibration. However, the
same procedures apply to the
One-Point
calibration option also. The user would
simply follow the same procedures and ignore references to the second point.
Measuring Calibration Standards – For the First Time
Note:
It is assumed that the temperature sensor is connected to the MiniMax,
and semi automatic or automatic mode has been selected. The
temperature sensor should also be attached to the specific calibration bar
being measured. If an external pyrometer or other method will be used, the
temperature mode should be set to manual, and the temperature manually
entered into the gauge.
1) With the menu items still active from the previous section, press the
key
multiple times to tab right and the
key multiple times to tab left until the
CAL
menu is highlighted and displaying the submenu items.
2) Press the
and
arrow keys multiple times to scroll through the sub
menu items until
ZERO MODE
is highlighted.
3) Press the
and
arrow keys multiple times to scroll the zero mode
options until the
TWO POINT
option is displayed. Each time the arrow keys
are press, and confirmation screen will appear letting the user know that the
ZERO MODE
is being changed. Pressing the
key each time will allow
the mode to be changed.