Installation
22
Operator’s Manual
•
Setting for SatSeconds is ON
•
Experiencing no alarm condition
•
Operating at ambient temperature of 25°C (±5°C)
Note:
Two types of battery are available: the standard 5- hour and optional 10-hour.
Note:
Even if the monitoring system is turned off, the
Battery Charge Indicator
remains lit
while the battery recharges.
Note:
A full charge of a depleted battery takes more than four (4) hours for a 5-hour battery
or eight (8) hours for a 10-hour battery.
Plug the monitoring system into an AC outlet to charge the battery for a minimum of
three (3) minutes prior to turning on any monitoring system with a completely
discharged battery. When operating on internal battery, the monitoring system
battery status icon indicates the battery charge condition.
To charge the internal battery:
1.
Connect the monitoring system to AC power to charge a low or depleted battery.
Reference
2.
Verify the
Battery Charge Indicator
lights.
3.4.3
Connecting a Nellcor™ Pulse Oximetry Sensor
WARNING:
Incorrect application or use of an SpO
2
sensor can cause tissue damage. Do
not wrap the sensor too tightly, apply supplemental tape, or leave a sensor
too long on one place. Inspect the sensor site as directed in the
Directions for
Use
to ensure skin integrity, correct positioning, and adhesion of the sensor.
WARNING:
Do not use any other cables to extend the length of the Covidien-approved
interface cable. Increasing the length will degrade signal quality and may lead
to inaccurate measurements.
Summary of Contents for Nellcor
Page 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM ...
Page 8: ...viii ...
Page 10: ...Page Left Intentionally Blank x ...
Page 18: ...Introduction 8 Operator s Manual Page Left Intentionally Blank ...
Page 100: ...Page Left Intentionally Blank 90 Operator s Manual Theory of Operations ...
Page 119: ......