Product Specifications
102
Operator’s Manual
Table 11-11.
Patient Applied and Patient Isolation Risk Current
Patient Applied Risk Current
Condition
AC Line Polarity
Neutral Line
Power Line
Ground Cable
IEC 60601-1
ANSI/AAMI
ES 60601-1: 2005
Normal
Normal
Closed
Closed
100 µA
Single Fault
Open
Closed
500 µA
Closed
Open
Normal
Reversed
Closed
Closed
100 µA
Single Fault
Open
Closed
500 µA
Closed
Open
Patient Isolation Risk Current
Condition
AC Line Polarity
Neutral Line
Power Line
Ground Cable
IEC 60601-1
ANSI/AAMI
ES 60601-1: 2005
Single Fault
Normal
Closed
Closed
5000 µA
Reversed
Closed
Closed