External Data Communication
Operator’s Manual
47
Figure 5-2.
Graphical Trend Data Screen
To scroll through Graphical Trend Data:
1.
Rotate the jog dial to highlight
Scroll
.
2.
Press the jog dial to activate scrolling.
3.
Rotate the jog dial to scroll through the trend data.
•
Clockwise
rotation moves forward to newer data.
•
Counterclockwise
rotation moves backward to older data.
4.
After reviewing trend data, press the RETURN button to exit the graphical trend
view.
5.4
External Data Communication
WARNING:
Any connections between this monitoring system and other devices must
comply with applicable medical systems safety standards such as IEC 60601-1.
Failure to do so may result in unsafe leakage current and grounding
conditions.
The monitoring system provides external connectors on the right and rear
panels to support data communication.
•
Nurse call interface (RJ11) —
Allows caregivers to remotely monitor patient
alarms via the nurse call system of the institution.