![background image](http://html1.mh-extra.com/html/covidien/nellcor/nellcor_operators-manual_2673894111.webp)
Manufacturer’s Declaration
Operator’s Manual
101
Table 11-10.
Earth and Enclosure Leakage Current Specifications
Earth Leakage Current
Condition
AC Line Polarity
Line Cord
Neutral
Line Cord
IEC 60601-1
IEC 60601-1
ANSI/AAMI
ES 60601-1: 2005
Normal
Normal
Closed
Closed
500 µA
300 µA
Single Fault
Open
Closed
1000 µA
Closed
Open
Normal
Reversed
Closed
Closed
500 µA
300 µA
Single Fault
Open
Closed
1000 µA
Closed
Open
Enclosure Leakage Current
Condition
AC Line Polarity
Neutral
Line Cord
Power Line
Ground
IEC 60601-1
ANSI/AAMI ES 60601-1: 2005
Normal
Normal
Closed
Closed
100 µA
Single Fault
Open
Closed
500 µA
Closed
Open
Normal
Reversed
Closed
Closed
100 µA
Single Fault
Open
Closed
500 µA
Closed
Open