![background image](http://html1.mh-extra.com/html/covidien/nellcor/nellcor_operators-manual_2673894040.webp)
Operation
30
Operator’s Manual
4.4.2
QUICK ACCESS Menus
For quick access to alarm limit settings, use the menu options listed here.
1.
SpO
2
Menu —
Provides access to SpO
2
alarm limit settings, alarm inhibition, and
SatSeconds alarm management option. Reference
for basic information. The adult default setting for the SatSeconds alarm manage-
ment option is 100. Other options include OFF, 10, 25, and 50. Reference
onds™ Alarm Management Feature
Figure 4-3.
QUICK ACCESS SpO
2
Menu with Audio Alarm Selected
SATSECONDS option
SatSeconds™ alarm management setting
(Off, 10, 25, 50, 100)
100
PATIENT MODE Menu
ADULT option
Sets Alarm Limits to standard default thresholds for adult
patients
PEDIATRIC option
Sets Alarm Limits to standard default thresholds for pediatric
patients
NEONATE option
Sets Alarm Limits to standard default thresholds for neonate
patients
SpO
2
WAVEFORM Menu
SWEEP SPEED option
6.25 mm/s, 12.5 mm/s, 25.0 mm/s
25.0 mm/s
TABULAR TREND option
Tabular trend view of trend data
--
GRAPHICAL TREND option
Graphical trend view of trend data
Table 4-1.
Menu structure and available options (Continued)
Item
Available Selections
Default
Summary of Contents for Nellcor
Page 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM ...
Page 8: ...viii ...
Page 10: ...Page Left Intentionally Blank x ...
Page 18: ...Introduction 8 Operator s Manual Page Left Intentionally Blank ...
Page 100: ...Page Left Intentionally Blank 90 Operator s Manual Theory of Operations ...
Page 119: ......