Operation
42
Operator’s Manual
To pause an audible alarm:
1.
Press the
Alarm Audio Paused
button to immediately pause the alarm tone. The
alarm resumes after the
Alarm Audio Paused
period, if the alarm condition
remains.
2.
Take the appropriate corrective action.
Note:
Press the Alarm
Audio Paused
button to cancel audible alarms caused by technical
errors. Audible alarms for battery failure and physiological alarms cannot be canceled
without appropriate corrective action.
To re-enable the audio tones during the
Alarm Audio Paused
period, press
the Alarm
Audio Paused
button again. If another alarm occurs during the
Alarm Audio Paused
period, the monitoring system re-enables all audio
tones.
To pause an audible alarm:
1.
Press the
Alarm Audio Paused
button.
2.
To cancel, press the
Alarm Audio Paused
button again.
If the
Alarm Audio Paused
period is enabled, the audible alarm is not active
for the specified time interval and the
Alarm Audio Paused
icon appears
above the appropriate alarm limit icon.
Note:
To disable limit violation alarms, use the Alarm Limits menus. Reference
Table 4-3.
Audio Status
Alarm Icon
Status
Alarm active
Alarm audio paused
Alarm audio OFF