Product Specifications
100
Operator’s Manual
11.9.2
Sensor and Cable Compliance
WARNING:
The use of accessories, sensors, and cables other than those specified may
result in inaccurate readings of the monitoring system and increased emission
of the monitoring system.
11.9.3
Safety Tests
Ground Integrity
100 milliohms or less
Leakage Current
The following tables display the maximum earth and enclosure leakage current
allowed, as well as patient leakage.
Table 11-9.
Cables and Sensors
Item
Maximum Length
Sensors
Pulse oximetry sensor cable
1.6 ft. (0.5 m)
Cables
Power cable
10.0 ft. (3.0 m)
Nurse call cable
5.9 ft. (1.8 m)
Pulse oximetry interface cable
10.0 ft. (3.0 m)