![background image](http://html1.mh-extra.com/html/covidien/nellcor/nellcor_operators-manual_2673894114.webp)
Clinical Studies
104
Operator’s Manual
A.3
Study Population
A.4
Study Results
Accuracy was calculated using the root mean square difference (RMSD).
Table A-1.
Demographic Data
Type
Class
Total
Gender
Male
5
Female
6
Race
Caucasian
8
Hispanic
2
African American
1
Asian
0
Age
--
19-48
Weight
--
108-250
Skin pigment
Very light
2
Olive
5
Dark olive/Medium black
3
Extremely dark/Blue black
1
Table A-2.
SpO
2
Accuracy for Nellcor™ Sensors vs. CO-oximeters
SpO
2
Decade
MAX-A
MAX-N
MAX-FAST
Data
Points
Arms Data
Points
Arms Data
Points
Arms
60-70
71
3.05
71
2.89
71
2.22
70-80
55
2.35
55
2.32
55
1.28
80-90
48
1.84
48
1.73
48
1.48
90-100
117
1.23
117
1.68
117
0.98
Summary of Contents for Nellcor