Manufacturer’s Declaration
Operator’s Manual
99
For transmitters rated at a maximum output power not listed above, estimate
the separation distance (
d
) using the equation in the corresponding column,
where
P
is the maximum output [power rating of the transmitter in watts (
W
)]
according to the transmitter manufacturer.
Note:
Portable and mobile RF communications equipment can affect medical electrical
equipment. Such RF equipment should be used no closer to any part of the monitoring
system, including cables, than the recommended separation distance calculated from
the equation appropriate for the frequency of the transmitter.
Table 11-8.
Recommended Separation Distances
Immunity
Test
IEC/EN 60601-1-2
Test Level
Compliance
Level
Electromagnetic Environ-
ment Guidance
Frequency of
Transmitter
Equation for Separation
Distance (
d
)
Conducted RF
IEC/EN 61000-4-6
3 Vrms
150 kHz
80 MHz
3 Vrms
150 kHz
80 MHz
150 kHz to 80 MHz
Radiated RF
IEC/EN 61000-4-3
3 V/m
80 MHz
800 MHz
3 V/m
80 MHz
800 MHz
80 MHz to 800 MHz
3 V/m
800 MHz
2.5 GHz
3 V/m
800 MHz
2.5 GHz
800 MHz to 2.5 GHz
Rated Maximum
Output Power (
P
) of
Transmitter in Watts
Separation Distance in Meters
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.10
0.38
0.38
0.73
1.00
1.20
1.20
2.30
10.00
3.80
3.80
7.30
100.00
12.00
12.00
23.00
d
1.2
P
=
d
1.2
P
=
d
2.3
P
=
Summary of Contents for Nellcor
Page 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM ...
Page 8: ...viii ...
Page 10: ...Page Left Intentionally Blank x ...
Page 18: ...Introduction 8 Operator s Manual Page Left Intentionally Blank ...
Page 100: ...Page Left Intentionally Blank 90 Operator s Manual Theory of Operations ...
Page 119: ......