2-5
(1) HANDSWITCHED MONOPOLAR
ACTIVE JACK - Accepts industry-standard
handswitching monopolar active accessories.
Output is delivered only when handswitch con-
trols are operated. See below for connections.
(2) FOOTSWITCHED MONOPOLAR
ACTIVE JACK - Accepts Bovie #12 monopolar
active accessories and adapters. Output is deliv-
ered only when foot switch is operated.
(3) BIPOLAR OUTPUT - Accepts Conmed
bipolar plugs and bipolar cords with 1/4" banana
plugs. The small hand switch pin is unused.
Output is delivered only in BIPOLAR COAG
mode when coag footswitch is depressed.
(4) RETURN ELECTRODE JACK - Accepts
Industry standard ARM
TM
2-pin dispersive elec-
trode cord plugs. Proper connection is required
for monopolar operation.
(5) POWER SWITCH - Controls AC power on
(1) and off (O).
(6) RETURN ELECTRODE ADAPTER -
Provides for use of CONMED Bioplus single full
dispersive electrodes when connected to return
electrode jack (4).
HAND CONTROL
FOOT CONTROL
BIPOLAR
F
I
O
1
2
3
4
5
6
FIGURE 2.2 OUTPUT PANEL
CUT
COAG
RF
HANDSWITCHED ACTIVE CONNECTIONS