A-1
APPENDIX A
HLP CODES AND POSSIBLE CAUSES
See Section 4.5.1 and 3.5.1 for explanation:
Meaning (dependent
HLP
on when performed
Code
as noted)
When Performed
Possible Causes
1
Read/Write ability
Initialization
-Faulty A3U3(8031)
of 8031 internal
and operation
memory is impaired.
2
Program memory CRC
Initialization
-Faulty A3U1 (27C64)
check error.
and operation
-Faulty A3U3 (8031)
-Open or shorted Address, Control,
or Data bus lines.
3
Calibration memory
Initialization
-Calibration incomplete when
CRC check error.
and operation
exited CAL mode.
-Faulty A3U2 (X2001)
-Faulty A3U3 (8031)
-Open or shorted address, control, or
data bus lines.
-Faulty A3U6 (74LS00) or A3U8
(74LS32)
4
Watchdog Timer
Initialization
-WDT timing parts incorrect
pulse width is
-Microprocessor frequency incorrect
incorrect
-Faulty A3U7 (4538)
-Faulty A3U3 (8031)
-Faulty A3U6 (74LS00)
-Faulty A3U10 (74LS00)
4
Watchdog Timer
Operation
-Faulty A3U3 (8031)
detected WDTSTB
-Faulty A3U1 (27C64)
outside of timing
-Processor 10 Mhz clock frequency
window.
incorrect
5
RF PA supply current
Initialization
-Incorrect base or gate drive
is out of range for test
-Shorted PA power transistor
drive setting.
-A4D2 or RV1 shorted
-A3J5 or A4P3 disconnected
-IFAIL or VSENSE circuits faulty
-Open fuses(s) or transistor(s) in PA
-Bipolar relay stuck closed
5
RF PA supply current
Operation
-Incorrect base or gate drive
is too high for power
-Shorted PA power transistor
setting.
-A4D2 or RV1 shorted
-VSENSE circuit faulty (cut modes)
-Bipolar relay stuck closed
Summary of Contents for sabre 180
Page 1: ...Service Manual...
Page 6: ...This page intentionally left blank...