1-6
OUTPUT ACCURACY:
•Setting: +/- 10%, +/- 5W to rated load @ rated mains.
•Mains: < 5% over mains voltage range.
•Power Decrease: < 5%, 50W output, 20 sec.
OUTPUT ADJUSTMENT:
•Settings: by digital knob from 0 to maximum in 1 watt steps.
•Display: seven segment digital display of watts to rated load resolved to 1 watt.
ACTIVATION CHARACTERISTICS:
•Activation tone volume adjustable 40-65 dbA via rear panel knob.
•Independence: Only the activated accessory will be live. Only one accessory may be activated at a time.
•Control switch activation resistance: < 1000 ohms.
ACCESSORY CONNECTIONS:
•CONMED makes available adapters to permit use of accessories with other than standard connectors:
•Monopolar Active: 3-pin handswitched and Bovie #12 footswitched.
•Monopolar Return: 2-pin A.R.M./REM capable.
•Bipolar: Two 1/4” (6.4mm) banana jacks.
•Footswitch: 4-pin MS connector for Aspen dual-treadle footswitch.
ALARMS:
•Return Electrode: 1000 Hz, > 65 dbA
•Single Foil acceptance: 0-10 ohms nominal
•Dual Foil acceptance: 10-150 ohms nominal
•Dual Foil alarm threshold: monitor set resi20% nominal
•Control error: 1710 Hz pulsating, 40-65 dbA
•Machine fault: 1000 Hz, 65 dbA continuous
DIMENSIONS:
•Size: 6" H (153mm) x 12" W (305mm) x 16" D (407mm)
•Weight: 20 lbs. (9.1 kg)
Note: Allow clearance of 2" (50mm) on each side and 3" (75mm) above for cooling.
SHIPPING & STORAGE:
Prior to shipment or storage, unit should be enclosed and sealed in polyethylene bag
and placed in original carton using original packing materials.
SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE
MODE
ACTIVATION MEANS
ACTIVATION TONE
COAG MODE
MONO CUT
HAND OR FOOT
520 Hz
EITHER
MONO COAG
HAND OR FOOT
250 Hz
MONOPOLAR
BIPOLAR COAG
FOOT ONLY
250 Hz
BIPOLAR