A-2
Meaning (dependent
HLP
on when performed
Code
as noted)
When Performed
Possible Causes
6
Attempted entry
Initialization
-RF shield not in place
into a guarded
and operation
-Faulty A3U3 (8031)
memory location or
-Faulty A3U1 (27C64)
CPU failure.
7
Watchdog Timer
Initialization
-Faulty A3U5
hardware does not
-Faulty A3U13
disable RF drive.
-Faulty A3U23
-+12 supply on A3 incorrect
-Faulty A3R50 or R51
8
Detection of a
Initialization
-Shorted or pressed hand or foot control
shorted hand or
-Continuity detector failure
foot control
-A3U11 (8255) failure
-Shorted bypass caps on control lines
9
Attempted entry to
Cal Mode
-A3S1 set in Cal position
Cal mode without
-Faulty/Disconnected control panel
“PAD” button pressed.
-Faulty A2U1
10
Incomplete Cal
Cal mode
-One or more Cal points not
calibrated on return to CAL menu.
11
Non-monotonic
Cal mode
-OPERATOR error when calibrating
Calibration
power
-Faulty Waveform Generator
-Faulty Base Voltage Generator
-Faulty A3U2 (X2001)
12
The power setting in
Operation
-RF shield not in place
data memory does not
-Faulty A2U1
match that displayed
-Faulty A3U3
for this mode.
-Faulty bus connection to A2J2
13
Control Panel
Initialization
-Pushbutton shorted in control panel
button stuck
-A2U1 or RN1 defective
-Operator error
21
A.R.M. voltage
Operation
-Shorted VARM sense line
Less than +0.4V
-Shorted A3VR1
-Shorted A4Q5 or Q6
-Shorted A4T2