1.2 SPECIFICATIONS
MAINS INPUT REQUIREMENTS:
•Single Phase AC / 48-62 Hz / 200W
•Power Cord: 3 conductor, #18 AWG (1.5mm2) Cu, CEE22 250V 6A Mains Connector
MAINS FREQUENCY LEAKAGE:
•Chassis: <100 uA
IEC Class 1, Type CF, Defibrillator Proof
•Patient: <10 uA
RF OUTPUT SPECIFICATIONS:
RF LEAKAGE: <100 mA, RF ISOLATED
DUTY CYCLE: 15s ON, 30s IDLE
COOLING: Natural convection, no fan
1-5
•The heat dissipation capability of the heat sink is
severely impaired by activating the Sabre 180 in
other than its normal operating position. There
are no tests requiring operation in any other posi-
tion.
•Consult the factory for advice before making
any modifications to the unit.
•Ensure that the RUN-RST-CAL switch is set to
RUN and that the two top cover screws are tight-
ened before returning the unit to service.
MAX. OPEN
MAX.
RATED
CARRIER
CREST FACTOR
CIRCUIT
PULSE
POWER
LOAD
FREQ.
50% POWER
VOLTAGE REPETITION
MODE
WATTS
OHMS
KHz
RATED LOAD
VPP
FREQ. KHz
CUT PURE
100
500
417
1.8
1600
CONTIN.
CUT BLEND 1
65
500
417
2.3
1650
20 KHz
CUT BLEND 2
50
500
417
3.5
1750
20 KHz
MONO. COAG
80
500
540
7.0
6500
33 KHz
BIPOLAR COAG
50
50
1050
2.4
400
20 KHz
MAINS VOLTAGE
CURRENT
VRMS
ARMS
CATALOG NO.
RATED
MIN
MAX
MAX
IDLE
60-5800-001
120
104
127
2.4
0.5
60-5800-002
100
90
110
2.9
0.6
60-5800-003
230
208
254
1.3
0.3
F