Disconnect the connector corresponding to the temperature probe you want to calibrate
from the
electronic power board
34
:
disconnect connector CN12 for sensor PT1,
connector CN13 for sensor PT2 or connector CN14 for sensor PT3.
Using the cable included in the kit, connect connector CN1 of the calibration tool (which
corresponds to a 1000 Ohm sample coil, corresponding to 0°C) to connector CN12, CN13
or CN14 of the
power board
34
:
Wait for at least 10 seconds and press ENTER.
Technician Menu