32
buzzer will sound for 3 seconds when the
“Maximum water level”
is reached to indicate
that the tank is full. If you do not hear the signal, excess water will flow out from the filler.
•
If the alarm is displayed during the cycle
, fill the tank and proceed as shown in the
INSUFFICIENT STEAM ALARM procedure. Furthermore, check and replace the
clean
water tank
10
level probe insulation.
Note:
In 18 S Dynamica and 18/24 S - B - BHD models, the steriliser introduces water into the
chamber by means of the pump to generate steam until the water touches the
chamber
probe
21
. If water is not sufficient during the water inlet phase, it means that there is
leakage of steam from the valves, from the chamber or from the pipes.
•
In extreme conditions, when a SPECIAL cycle is run with three fractioned vacuum pulses
and with a very heavy porous load, the expected minimum level may not be sufficient.
Reduce the load or use two vacuum pulses only.
Note:
Steam is generated in 6 S Dynamica by introducing water in the chamber using the
pump and measuring the required amount by means of the volumetric counter. If there
are no leakages of water from underneath the steriliser, check correct operation of the
volumetric counter and replace it, if required.
Chamber level alarm
This alarm indicates that the time required to fill
chamber
09
with the required amount of
water has elapsed.
In 18 S Dynamica and 18/24 S - B - BHD models this alarm is displayed when chamber probe
21
does not detect water in the chamber. In 6 S Dynamica the alarm appears when the
volumetric counter does not detect the required amount of water before the timeout.
•
When using 18 S Dynamica and 18/24 S - B - BHD, check inclination of the steriliser (see
USE AND MAINTENANCE MANUAL, chapter INSTALLATION).
•
If the problem persists, check:
- that the
water inlet pump
35
and the solenoid valve
EV1
are pumping water into
chamber
09
;
- if solenoid valve
EV1
winding has blown;
- the
water inlet pump
35
;
- the water filter;
-
fuse
F1
on the
power board
34
.
If none of these causes are found, check
chamber probe
21
which is probably not
detecting the level and the probe electric connection.
To check correct operation probe:
- open
door
16
and remove the entire contents of
chamber
09
;
- remove
tray support
08
;
- remove
coil grid
19
by loosening the corresponding screw;
- reassemble back to front, i.e. by turning the grid and seal horizontally;
- switch the steriliser on in STAND-BY mode;
- run Chamber load test from the Technician Menu (option 15).
The probe could be dirty or the power cable could be interrupted. Furthermore, check
level probe grounding.
Alarms
Summary of Contents for STERILCLAVE 18 B
Page 2: ...2...
Page 6: ...6 Graphical reference representation Model 6 S Dynamica Contents...
Page 7: ...7 Graphical reference representation Model 18 Contents...
Page 8: ...8 Graphical reference representation Model 24 Contents...
Page 9: ...9 Graphical reference representation of common components 40 Contents...
Page 10: ...10...
Page 63: ...63 Diagrams and wirings...
Page 75: ...75...