30
Cooling pause
This message indicates that the temperature in
chamber
09
is too high to start or continue
a cycle (see USE AND MAINTENANCE MANUAL, chapter OPERATION).
The steriliser will display this message in STAND BY mode or after starting a cycle and will
stop only during the pre-vacuum phase. This message normally appears when several
cycles are run one after the other and no time is allowed for the chamber to cool down.
•
In such conditions, simply let the chamber cool down. To speed up cooling, open
door
16
, otherwise cooling will take considerably longer. To open the door, press STOP,
cancel the MANUAL STOP ALARM and wait for a few seconds for the pressure inside
the chamber to equalise ambient pressure to allow the door to be opened.
•
With the door closed
the operating cycle will resume automatically when the allowed
temperature is reached.
•
If you opened the door
, start the cycle again after cooling down the steriliser.